Long-Term Care Bed • D500 • 5 Function Ultra Low Bed (Reorder No. 12005)

Long-Term Care Bed • D500 • 5 Function Ultra Low Bed (Reorder No. 12005) 

11

EQUIPMENT OPERATION

1.  POSITIONING THE BED

 

To position the bed unlock the casters. Once the bed 

is in its desired location, lock all casters in place.

2.  USING THE HAND PENDANT

 

The UP arrow (▲) indicates raising the corresponding 

parts of the bed. 

 

The DOWN arrow (▼) indicates lowering the corre-

sponding parts of the bed. 

Use button 1 and 2 to 

adjust the backrest up or 

down. 
Use button 3 and 4 to 

adjust the kneebreak up 

or down.
Use button 5 and 6 to 

adjust the backrest and 

kneebreak up or down 

simultaneously. (Auto 

Contour feature)
Use button 7 and 8 to 

raise or lower the bed 

platform.
Use button 9 and 10 

to tilt the bed into re-

versed Trendelengburg 

and Trendelenburg posi-

tions.

2.1  ENTRAPMENT SAFETY MEASURE

 

The bed is equipped with an Entrapment Safety 

Measure. If the bed platform is higher than 12.5" from 

the fl oor, long press button 8 will lower the bed to 

12.5". To continue lowering the bed, release and long 

press button 8 again. Ensure no body parts or 

obstacles are within the bed’s moving range. Ensure 

the bed rails/bars are raised. 

2.2  RESETTING THE BED

 

Where a connection fault is identifi ed and rectifi ed, 

the control system will require a reset in order to 

operate again. If the fault occurred on the motors for 

platform profi ling, lower the corresponding motor to 

the lowest position for reset; where the fault occurred 

on the motors for bed height adjustments, raise the 

bed to the highest position for reset. 

EQUIPMENT OPERATION 

(Continued)

3.  ADJUSTING THE EXTENDED 

HEEL LIFT

 

The end foot section can be adjusted manually to 

further alter the kneebreak angle.

 

1.  Grip the end section of the bed platform. Lift up 

slowly to the preferred height. Ensure there is an 

audible click indicating that the deck is locked in 

place within the Ratcheting Bracket.

 

2.  To lower the end deck to its original position, lift 

up to the highest position and slowly lay the bed 

platform back down. 

  

4.  FOLDING AND TRANSPORTING 

THE BED

 

The bed can be disassembled and transferred in a 

compact manner using the transport bars.

 

1.  Reverse the bed installation procedure to disas-

semble the bed into four sections, the headboard, 

the footboard, the upper and lower bed decks. 

Keep the screws and washers.

 

2.  Position the transport bar with the deck connection 

brackets facing upwards. Insert a transport bar on 

each of the headboard connection brackets.

 

3.  Insert the footboard into the other side of the 

transport bar.

 

4.  The folded bed can be transported with the 

casters on the head and footboard. Reverse the 

procedure to reassemble the bed. 

  

Содержание 12005

Страница 1: ...Manufactured for Dynarex Corporation 10 Glenshaw Street Orangeburg NY 10962 USA www dynarex com Made in China Long Term Care Bed D500 5 Function Ultra Low Bed User Manual Reorder No 12005...

Страница 2: ...unlocked unintended movement may occur which may cause dam age to the resident or property damage Moving the bed while the caster locks are engaged may damage the bed or the floor Ensure all the cast...

Страница 3: ...dant Clip 1 M10 Hex Bolt 16 Transport Bar 2 Headboard 1 OPERATING RANGE Bed Height 2 5 25 6 Back Rest Angle 0 70 Knee Break Angle 0 27 Additional Manual Leg Lift Yes various positions Trendelenburg Re...

Страница 4: ...two socket screws on each side 4 Tilt the assembly back on its foot 5 Repeat the same procedure for foot board installation 2 3 ASSEMBLING THE BED DECK 1 With both the head and foot end assemblies st...

Страница 5: ...stalling the backup battery 2 Affix the backup battery with nuts and bolts to the left of the control box on the same holding plate 3 Open the control box locking lid by pressing in the sides and lift...

Страница 6: ...identified and rectified the control system will require a reset in order to operate again If the fault occurred on the motors for platform profiling lower the corresponding motor to the lowest positi...

Страница 7: ...ndant buttons are pressed Adjustment may be at the maximum or minimum position Power cord not connected or damaged Hand pendant or actuator connectors are loose Load is too heavy Control system has no...

Страница 8: ...roducts and our assurance to the consumer of years of dependable service The Warranty shall not apply under the following conditions Problems arising from normal wear Problems arising from failure to...

Страница 9: ...16 Long Term Care Bed D500 5 Function Ultra Low Bed Reorder No 12005 SERVICE RECORD DATE PERFORMED BY CONDITION REPORT...

Отзывы: