background image

12 

Long-Term Care Bed • D500 • 5 Function Ultra Low Bed (Reorder No. 12005)

Long-Term Care Bed • D500 • 5 Function Ultra Low Bed (Reorder No. 12005) 

13

EQUIPMENT OPERATION 

(Continued)

5.  OPERATING THE BED RAILS

 

METAL SWING RAIL

(REORDER NUMBER 12950)

 

Raise Bed Rail:

 Pull the bed rail upwards until the 

rail is securely locked in place. 

 

Lower Bed Rail: 

Squeeze the release lever at the 

bottom corner of the bed rail and slowly lower the bed 

rail to its lowest position.

 

COMPOSITE SWING RAIL

(REORDER NUMBER 12951)

 

Raise Bed Rail:

 Pull the bed rail upwards until the 

rail is securely locked in place.

 

Lower Bed Rail: 

Press the release button and slowly 

lower the bed rail to its lowest position.

 

Q ASSIST BAR

(REORDER NUMBER 12952)

 

Parallel Q Assist Bar Alignment:

 Lift up and rotate 

the bed rail until it is parallel to the bed edge and 

release the bed rail. Ensure it is securely locked in 

place.

 

Perpendicular Q Assist Bar Alignment: 

Lift up and 

rotate the bed rail until it is perpendicular to the bed 

edge, then release the bed rail. Ensure it is securely 

locked in place.

 

PIVOT ASSIST BAR

(REORDER NUMBER 12953)

 

Raise Bed Rail: 

Pull the bed rail upwards until the 

rail is securely locked in place.

 

Lower Bed Rail: 

Pull the release knob and slowly 

lower the bed rail to its lowest position.

TROUBLESHOOTING GUIDE

SYMPTOMS

POSSIBLE FAULTS

POSSIBLE SOLUTIONS

Bed idle when hand pendant 

buttons are pressed.

Adjustment may be at the maximum 

or minimum position.

Power cord not connected or 

damaged.

Hand pendant or actuator connectors 

are loose.

Load is too heavy.

Control system has not been reset 

after faults were fi xed.

Check if other buttons are working. 

Ensure the power cable is properly 

connected to the electrical socket 

and the bed.

Ensure tight connection of all 

connectors to control box.

Ensure the load weight is within the 

designated capacity. Reduce the 

load.

Reset the bed (see page 10).

Adjustment via hand pendant or Staff 

Control only partially occurs and 

stops.

The bed may be overloaded.

The bed may be constrained 

from moving.

Actuators connection may be 

loose or damaged. 

Thermal shut down may have 

occurred. 

If battery backup is installed, 

the battery may be low.

Ensure the load weight is within the 

designated capacity. Reduce the 

load.

Check for obstructions.

Ensure actuator is properly secured 

to the bed. 

Wait at least 20 minutes before 

trying the functions again. The bed is 

intended to be used for 2 min every 

20 min.

Plug bed into appropriate power 

source.

Bed not responding to hand pendant 

or Staff Control.

The lockout function may be en-

gaged.

Cables may be loose.

Thermal shut down may have 

occurred.

Ensure that the lockout function is not 

engaged on the Staff Control.

Ensure all cables are fully inserted.

Ensure the bed is plugged into 

appropriate outlet.

Unplug bed for 60 seconds and 

plug back in.

Wait at least 20 minutes before 

trying the functions again. The bed is 

intended to be used for 2 min every 

20 min.

Содержание 12005

Страница 1: ...Manufactured for Dynarex Corporation 10 Glenshaw Street Orangeburg NY 10962 USA www dynarex com Made in China Long Term Care Bed D500 5 Function Ultra Low Bed User Manual Reorder No 12005...

Страница 2: ...unlocked unintended movement may occur which may cause dam age to the resident or property damage Moving the bed while the caster locks are engaged may damage the bed or the floor Ensure all the cast...

Страница 3: ...dant Clip 1 M10 Hex Bolt 16 Transport Bar 2 Headboard 1 OPERATING RANGE Bed Height 2 5 25 6 Back Rest Angle 0 70 Knee Break Angle 0 27 Additional Manual Leg Lift Yes various positions Trendelenburg Re...

Страница 4: ...two socket screws on each side 4 Tilt the assembly back on its foot 5 Repeat the same procedure for foot board installation 2 3 ASSEMBLING THE BED DECK 1 With both the head and foot end assemblies st...

Страница 5: ...stalling the backup battery 2 Affix the backup battery with nuts and bolts to the left of the control box on the same holding plate 3 Open the control box locking lid by pressing in the sides and lift...

Страница 6: ...identified and rectified the control system will require a reset in order to operate again If the fault occurred on the motors for platform profiling lower the corresponding motor to the lowest positi...

Страница 7: ...ndant buttons are pressed Adjustment may be at the maximum or minimum position Power cord not connected or damaged Hand pendant or actuator connectors are loose Load is too heavy Control system has no...

Страница 8: ...roducts and our assurance to the consumer of years of dependable service The Warranty shall not apply under the following conditions Problems arising from normal wear Problems arising from failure to...

Страница 9: ...16 Long Term Care Bed D500 5 Function Ultra Low Bed Reorder No 12005 SERVICE RECORD DATE PERFORMED BY CONDITION REPORT...

Отзывы: