background image

Long-Term Care Bed • D500 • 5 Function Ultra Low Bed (Reorder No. 12005)

Long-Term Care Bed • D500 • 5 Function Ultra Low Bed (Reorder No. 12005) 

7

2.1  ASSEMBLING THE BED

 

The bed comes in with the sections (head board, foot 

board, upper bed deck and lower bed deck) stacked. 

Carefully lay each section on the fl oor.

2.2  ASSEMBLING THE HEADBOARD 

AND FOOTBOARD

 

1.  Position two long pieces of cardboard supports on 

the fl oor shoulder width apart.

 

2.  Carefully tilt the head board onto the cardboard 

supports. Position the plastic actuator casing at 

the lower center of the head board in between the 

two supports.

  

Cardboard 

supports

 

3.  Rotate the head end bed deck vertically. Aligned 

and slot the bed deck onto the head board. Secure 

the assembly with two socket screws on each 

side.

 

4.  Tilt the assembly back on its foot.

 

5.  Repeat the same procedure for foot board 

installation.

  

2.3  ASSEMBLING THE BED DECK

 

1.  With both the head and foot end assemblies 

standing on their feet, align and join the two decks 

together.

 

2.  Secure the two decks together with two socket 

screws on each side.

  

ASSEMBLY INSTRUCTIONS 

(Continued)

ASSEMBLY INSTRUCTIONS 

(Continued)

2.4  INSTALLING THE MATTRESS 

RETAINING SYSTEM

 

1.  Snap the mattress retainer onto the bed frame 

and push it down fi rmly. The mattress retainers for 

the sides will be longer than the ones for the bed 

ends. Ensure the mattress retainer sits fl ushed 

with the edge of the bed.

 

2.  Repeat the procedure until all six mattress retain-

ers are installed. There should be one retainer on 

either side of the bed and two on both ends of the 

bed.

  

  

2.5  INSTALLING THE HAND 

PENDANT CLIP

 

Clip the hand pendant clip over the headboard and 

push it down fi rmly.

Содержание 12005

Страница 1: ...Manufactured for Dynarex Corporation 10 Glenshaw Street Orangeburg NY 10962 USA www dynarex com Made in China Long Term Care Bed D500 5 Function Ultra Low Bed User Manual Reorder No 12005...

Страница 2: ...unlocked unintended movement may occur which may cause dam age to the resident or property damage Moving the bed while the caster locks are engaged may damage the bed or the floor Ensure all the cast...

Страница 3: ...dant Clip 1 M10 Hex Bolt 16 Transport Bar 2 Headboard 1 OPERATING RANGE Bed Height 2 5 25 6 Back Rest Angle 0 70 Knee Break Angle 0 27 Additional Manual Leg Lift Yes various positions Trendelenburg Re...

Страница 4: ...two socket screws on each side 4 Tilt the assembly back on its foot 5 Repeat the same procedure for foot board installation 2 3 ASSEMBLING THE BED DECK 1 With both the head and foot end assemblies st...

Страница 5: ...stalling the backup battery 2 Affix the backup battery with nuts and bolts to the left of the control box on the same holding plate 3 Open the control box locking lid by pressing in the sides and lift...

Страница 6: ...identified and rectified the control system will require a reset in order to operate again If the fault occurred on the motors for platform profiling lower the corresponding motor to the lowest positi...

Страница 7: ...ndant buttons are pressed Adjustment may be at the maximum or minimum position Power cord not connected or damaged Hand pendant or actuator connectors are loose Load is too heavy Control system has no...

Страница 8: ...roducts and our assurance to the consumer of years of dependable service The Warranty shall not apply under the following conditions Problems arising from normal wear Problems arising from failure to...

Страница 9: ...16 Long Term Care Bed D500 5 Function Ultra Low Bed Reorder No 12005 SERVICE RECORD DATE PERFORMED BY CONDITION REPORT...

Отзывы: