BlueWave® QX4® User Guide
33
Table 2. Troubleshooting Chart for
BlueWave LED QX4
Problem
Possible Cause
Corrective Action
BlueWave QX4
does
not power up
Power cord not plugged
in or damaged
Check power connection and
condition at power supply
“Brick” and controller
.
No electrical power at
Receptacle
Test Receptacle for power.
BlueWave QX4
powers up but the
LED head does not
produce light
LED intensity adjustment
set to 0%
Increase LED intensity setting.
LED cycle time is set to 0
seconds
Increase cycle time setting.
Interface cable
connections loose or
damaged
Check connections and condition
of Interface Cable.
In variable mode, the
cycle count is set to 0
Adjust the cycle count to the
desired number.
LED head is not
connected to the correct
port/channel
Verify that the head is
connected to the desired
port/channel.
BlueWave QX4
is
operating normally
and the head
suddenly stops
producing light
Over-temperature
shutdown was triggered
Check if the front panel over-
temperature indicator is red. If
so, allow the head to cool and
follow the instructions on the
screen.
LED head provides
only low-intensity
light
LED intensity adjustment
set to minimum
Increase LED intensity setting.
Contaminated/dirty lens
optics
Clean the surface of the Lens.
Footswitch does
not function
Footswitch not
connected
Connect footswitch.
Footswitch defective
Activate unit using the front
control panel. Replace the
footswitch if the unit operates
from the front control panel.