DYACON UA-1 Manual
Modbus
Basic Configuration Messages
The following commands can be used to change the serial port parameters of the air sensor.
Address Register
Access Type
Range
Data Type
Description
103
104
Read/Write
1-247
Default: 1
16-bit Signed Int
Modbus slave address.
104
105
Read/Write
0 = 1200 bps
1 = 2400 bps
2 = 4800 bps
3 = 9600 bps
4 = 19200 bps
5 = 38400 bps
16-bit Signed Int
Baud rate.
Default: 4 (19200 bps)
105
106
Read/Write
0 = None
1 = Odd
2 = Even
Default: 0
16-bit Signed Int
Parity
* Refer to Power Saving section for more details.