Lighting 

All  areas  must  be  illuminated  evenly  and  sufficiently to ensure all operations included in this manual and must be 
without shadows, reflections, glare or eyestrain.  

 

4.6 

PUMP INSTALLATION  

 

To ensure perfect pump operation, house and place it according to the following conditions: 

  Allow sufficient space on the perimeter sides of the pump and make sure to keep the motor ventilation side free. 

  Make  sure  the  free  space  adjacent  to  the  pump  allows  easy  access  to  components  for  inspection  or 

maintenance and also allows access for suitable lifting equipment. 

  The pump is equipped with mounting points. It is necessary to ensure it locks onto a perfectly horizontal plane in 

order to avoid tilting in case of transportation by the system user. 

  Some models are already equipped with rubber vibration dampers installed at pump attachment points. Whenever the 

model has not been equipped, ensure installation of such equipment so as not to transmit vibration to the pump. 

  Ensure there is ventilation in the room, or inside the machine housing the pump and prevent air coming in from 

the outlet or cooling fans, which could cause discomfort to personnel.

 

 

WARNING 
Do  not  install  the  pump  in  any  area  with  dust  or  other  materials  that  could  clog  or  rapidly  cover 
cooling surfaces of the heat exchanger (where installed). 

 

4.7 

MOTOR INSTALLATION (ONLY LC.105WR)  

 

It  is  possible  to  install  any  type  of  electric  or  hydraulic  motor  that  has  the 
features  described  in  the  table  of  technical  data,  with  flange  and  shaft 
corresponding to M100/4 - B5 size as per standard IEC-72. 

 

WARNING 
Install  fan/coupling  assembly  on  the  motor  following  these 
instructions: 

  Fit the assembly on the motor shaft up to reaching the stated measure, 

  Tighten screw “A” to firmly fix the assembly to the shaft. 

 

 

4.8 

USER SYSTEM 

Make sure that no harmful substances contaminate the user system during installation.  

If  you  wish  the  system  to  maintain  vacuum  even  when  pump  is  stopped,  install  a  cutoff  valve  between 
pump and system. 

Make sure that no vibrations or stresses are transmitted to the pump connection. 

 

4.9 

CONNECTION  

 

 

 

HAZARD 
Pump connections should be performed by skilled and trained personnel only. 

 

4.9.1 

ASSEMBLING THE INTAKE FILTER (ONLY LB.5WR)  

 

For packaging and shipping purposes, the filter is supplied disassembled. For connection to the pump a special ring 
nut connector with O-ring seal has been provided, which guarantees a perfectly tight coupling.  For assembly, just 
move the filter towards the intake and screw the ring nut onto the connector. 

 

4.9.2 

INTAKE AND OUTLET CONNECTIONS  

 

User system connections (both intake and outlet) must be performed with pipes with a diameter equal to or greater 
than the pump suction inlet. The weight of pipes or any expansions must not burden the pump.  
It  is  advisable  to  make  the  final  connection  to  the  pump  using  flexible  pipe  or  fittings.  It  is  important  to  tighten  all 
pipes and couplings. Very long pipes or pipes with a diameter that is too small diminish pump performance. 

 

HAZARD 
Make sure that discharged gases are let out of the workplace. 

Содержание LB.40WR

Страница 1: ...TECHNOLOGY OPERATING AND MAINT ENANCE INSTRUCTIONS Translation of the original instructions VACUUM PUMPS WITH RECIRCULATING LUBRICATION SYSTEM LB 5WR LC 25WR LB 40WR LB 60WR LC 105WR www dvp it info...

Страница 2: ...entele to reach its objectives For this reason our customers are the focus of our attention and we promote the sending of suggestions and or proposals for improvements that could be applied to our pro...

Страница 3: ...AND HANDLING 9 4 4 STORAGE 9 4 5 ENVIRONMENTAL CONDITIONS 9 4 6 PUMP INSTALLATION 10 4 7 MOTOR INSTALLATION ONLY LC 105WR 10 4 8 USER SYSTEM 10 4 9 CONNECTION 10 4 9 1 ASSEMBLING THE INTAKE FILTER ON...

Страница 4: ...model and serial number year of manufacture date of purchase detail information regarding problems verified 1 3 METHOD OF CONSULTATION For improved understanding of the information provided in this ma...

Страница 5: ...other identification plaques D V P Vacuum Technology S r l disclaims any liability for damage to persons or property resulting from improper use of the pump from tampering with its safety apparatus o...

Страница 6: ...mps are designed specifically to work with clean air and inert gases in atmospheres with a high amount of water vapour Any other use is prohibited The Manufacturer is not liable for any damage to pers...

Страница 7: ...capacity m h 5 6 Final pressure Abs mbar hPa 10 Max inlet pressure for water vapour mbar hPa 30 Max water vapour pumping rate Kg h 0 11 Motor power kW 1 3 0 25 0 37 0 25 0 45 Nominal r p m n min 1400...

Страница 8: ...2 Max inlet pressure for water vapour mbar hPa 50 Max water vapour pumping rate Kg h 0 9 Motor power kW 1 3 0 75 0 75 0 90 0 90 Nominal r p m n min 1400 1700 Noise level UNI EN ISO 2151 K 3dB dB A 62...

Страница 9: ...x inlet pressure for water vapour mbar hPa 40 50 Max water vapour pumping rate Kg h 1 2 2 1 Motor power kW 3 1 1 1 35 1 5 1 8 Nominal r p m n min 1400 1700 1400 1700 Noise level UNI EN ISO 2151 K 3dB...

Страница 10: ...mbar hPa 2 Max inlet pressure for water vapour mbar hPa 40 Max water vapour pumping rate Kg h 2 9 Motor power kW 3 2 2 2 7 Nominal r p m n min 1400 1700 Noise level UNI EN ISO 2151 K 3dB dB A 68 70 W...

Страница 11: ...e with local regulations Do not burn or disperse package components in the environment 4 3 TRANSPORT AND HANDLING HAZARD All transportation lifting and handling operation must be performed by qualifie...

Страница 12: ...to install any type of electric or hydraulic motor that has the features described in the table of technical data with flange and shaft corresponding to M100 4 B5 size as per standard IEC 72 WARNING...

Страница 13: ...uate for the power absorbed by the pump absorption values are shown on the pump motor rating plate taking into account the environmental conditions of operation HAZARD Always ground the pump Always in...

Страница 14: ...ime with the suction inlet completely open WARNING It is advisable not to start the pump more than 12 times per hour to avoid excessive energy consumption and damage to the pump especially for model L...

Страница 15: ...n and or repair of any kind except for those listed in this manual Only trained or authorised personnel have the necessary expertise to perform tasks with the technique appropriate for intervention HA...

Страница 16: ...e for about 10 minutes first so oil will become thinner stop the pump and disconnect it from the mains undo the oil filler plug get a container large enough to hold all oil and open the oil drainage v...

Страница 17: ...l viewer 5503002 Intake filter cartridge 1801019 D V P Vacuum Technology S r l disclaims all responsibility for any deterioration of pump performance or for damages caused due to use of non original s...

Страница 18: ...e the air exhaust filter Bearings damaged Contact Service Department Solid particles in the oil Change oil Vanes worn out Contact Service Department C Pump is noisy Motor coupling damaged where presen...

Страница 19: ...appear to be faulty due to negligence and or carelessness during use failure to observe the equipment operating instructions lack of maintenance incorrect installation and or maintenance maintenance c...

Страница 20: ...D V P Vacuum Technology srl Via Rubizzano 627 40018 San Pietro in Casale BO Italy Tel 39 051 18897101 Fax 39 051 18897170 Email info dvp it Cod 8702037 01 01 2011 R 0 EN...

Отзывы: