background image

increase the life expectancy of the pump.  Air will 
remain stationary in the RAD AQUA for four 
minutes before moving to the desiccant where it waits 
another four minutes before entering the RAD7.  
While in the RAD AQUA, the stationary air may 
approach equilibrium with the water thus inhibiting 
further radon transfer from the water to the air.  It 
will be about 15 to 20 minutes before that parcel of 
air returns to the RAD AQUA.  We can therefore 
estimate that the response time of the system will be 
increased by about 20 minutes if the pump is set to 
AUTO 

Having the pump on AUTO would normally be 
associated with having the RAD7 cycle time of 30 
minutes or longer.  So an extra 15 to 20 minutes on 
the response time will not be excessive. 

2.2.5  RAD7 Mode 

In AUTO (the default) mode, the RAD7 will 
automatically switch from SNIFF mode to NORMAL 
mode a

ft

er three hours into the run.  

Th

is is to take 

advantage of the additional counts provided by the 
214-Po decays that will, by then, have approached 
equilibrium with the (steady) radon concentration in 
the measurement chamber. 

For slow, long-term measurements with long cycle 
times AUTO mode for the RAD7 is appropriate.  

Th

RAD7’s response time will be a couple of hours or so.  

Th

e RAD AQUA can be running with a low water 

fl

ow rate and the RAD7 pump can be on AUTO also. 

For fast response, however, it is essential to force the 
RAD7 to stay in SNIFF mode (

Setup, Mode, 

Sniff

 

[ENTER]

).  It will then always count only the 

218-Po decays, giving it a 13-minute, 95% response 
time. 

2.3  Long Term Measurement 

2.3.1  Desiccant 

As set up, above, the system will continue making 
measurements inde

fi

nitely. 

Th

ere are, however, 

various resources that are being used up in the 
process, and which must be replenished.  

Th

e most 

obvious is the desiccant.  A new, or regenerated, 
laboratory drying unit will normally last for about ten 
days of continuous use in a temperate climate.  In this 
application, however, it is receiving saturated air and, 
therefore, will be hydrated more quickly.  When the 
remaining length of blue (dry) desiccant is less than 
one inch, the desiccant should be replaced.  Please see 
the RAD7 manual on desiccant regeneration.  If the 
desiccant is not replaced, and the humidity in the 
instrument rises above about 20%, then the 
sensitivity drops o

 and the reading is lower than the 

true value. 

2.3.2  Memory 

Th

e capacity of the internal memory of the RAD7 is 

1,000 records or cycles.  If each cycle is half an hour, 
that would be data for 500 hours, or just over 20 days.  
Every time the desiccant is changed, therefore, all the 
stored data should be downloaded to a PC, backed 
up, and erased from the RAD7 memory.

Section 2 

 

Measurement Procedure

 11

Содержание RAD AQUA

Страница 1: ...RAD AQUA Continuous Radon in Water Accessory for the RAD7 User Manual...

Страница 2: ...rature Probe 8 1 3 Water Flow 8 1 3 1 Water Source 8 1 3 2 Water Level 8 1 4 Air Flow 8 1 4 1 Continuous Pumping 8 1 4 2 Pump on Auto 8 1 4 3 Mode set to Auto 8 1 5 Protocol 9 1 5 1 RAD7 Protocol 9 1...

Страница 3: ...Exchanger to RAD7 method 2 14 4 4 Speed of Response 15 5 DRYSTIK 16 5 1 Passive DRYSTIK 16 5 2 Active DRYSTIK 16 5 3 E ect on Response Time 16 5 4 Custom designed Active DRYSTIK 16 6 BOATING 18 6 1 Re...

Страница 4: ...ant Regeneration 20 7 5 RAD AQUA Troubleshooting 20 7 5 1 Rising Water Level 20 7 5 2 Spray Chamber Fills With Water 21 7 5 3 Air Path Integrity 21 7 5 4 Poor Spray Formation 21 7 5 5 Water Overflowin...

Страница 5: ...temperature the coe cient is about 0 25 That means there is four times higher concentration of radon in the air than in the water so there is in e ect a gain of four times in the sensitivity of the s...

Страница 6: ...plications where speed of response is a major goal WL1 is a compromise between the two 1 1 2 Temperature Probe The probe is inserted through the stem adapter A little petroleum jelly may help it to sl...

Страница 7: ...nt left at this end to the air inlet filter with 1 8 ID tubing at the filter end which is then placed on the RAD7 INLET The Luer taper ensures an airtight connection Fig 2 RAD7 AQUA Standard Setup 1 2...

Страница 8: ...ill be observed inside the body of the exchanger Water will accumulate inside the base and overflow out through the hose connector s If the flow rate is very high it may be necessary to utilize both h...

Страница 9: ...ntrations to gain better statistics choose AUTO SETUP THORON Choose OFF SETUP PUMP Choose ON or AUTO depending on choice see above SETUP TONE Choose what you like SETUP FORMAT Choose what you like but...

Страница 10: ...100 counts would give a reading with a standard deviation of 10 At very low concentrations it may take hours and averaging over many cycles to reach a su ciently precise value 2 2 2 In uencing Factor...

Страница 11: ...QUA can be running with a low water flow rate and the RAD7 pump can be on AUTO also For fast response however it is essential to force the RAD7 to stay in SNIFF mode Setup Mode Sniff ENTER It will the...

Страница 12: ...radon concentrations This water temperature data is obtained by exporting it from the temperature logger software see Section 3 1 5 below For specific instructions on downloading and graphing RAD7 dat...

Страница 13: ...emperature data is stored in the temperature logger It is necessary to download the RAD7 data to a PC preferably using DURRIDGE s CAPTURE software and to download the temperature logger data using the...

Страница 14: ...say 4 L min will mean a time delay for the thoron to reach the exchanger of no more than 2 5 seconds 4 3 2 Exchanger to RAD7 method 1 The RAD7 pump will typically generate an air flow rate of around 0...

Страница 15: ...ponse of the RAD7 to thoron is limited not be the half life of the polonium daughter but by the time it takes to get the sample into the measurement chamber We have seen that with the highest sensitiv...

Страница 16: ...7 will keep the air sample in the RAD7 below 7 RH It will also last for a very long time and add only a tiny volume to the air loop maintaining a fast response See Fig 3 5 3 E ect on Response Time Wit...

Страница 17: ...Fig 3 RAD AQUA con guration with Active DRYSTIK and Water Switch Section 5 DRYSTIK 17...

Страница 18: ...eded increases significantly The Rule 3700 delivers 6 L min through a W4 nozzle but it takes over 15A One 8 hour run would completely discharge a 120 AH battery 6 1 2 Minimizing T1 with reduced air vo...

Страница 19: ...reducing the advantage of the three pumps to only 5 to 10 minutes But with the Rule 3700 the extra 20L to 40L of water would take only 3 or 4 minutes to deliver thus preserving almost all the 20 minu...

Страница 20: ...ir sockets and push the lid onto its hinges The wrap will make the RAD7 almost watertight If it tears it can be easily replaced at any time The instrument should in any case be returned every year for...

Страница 21: ...or leaks 7 5 3 Air Path Integrity When drawing a sample from a remote location air path integrity is essential to prevent dilution of the sample with ambient air Always make sure that there are no lo...

Страница 22: ...ter inputs into river dominated areas important The Chao Phraya River Gulf of Thailand Limnology and Oceanography 51 2232 2247 Povinec P P P K Aggarwal A Aureli W C Burnett E A Kontar K M Kulkarni W S...

Страница 23: ...2008 Determination of transport rates in the Yellow River Bohai Sea mixing zone via natural geochemical tracers Continental Shelf Research 28 19 2700 2707 167 Povinec P P H Bokuniewicz W C Burnett J C...

Страница 24: ...67 9556 Fax 978 667 9557 Web www durridge com Email service durridge com Copyright 2018 DURRIDGE Company Inc All rights reserved DURRIDGE the DURRIDGE logo and the Works with the DURRIDGE logo are tra...

Отзывы: