Troubleshooting
Fault
Probable cause
Solution
X-ray image is stretched
lengthwise with bright,
horizontal stripes
Incorrect light protection cover or
image plate used
i
Only use original accessories.
X-ray image split vertically
into two halves
Dirt in the laser slit (e.g. hair, dust)
i
Clean the laser slit.
X-ray image with small
bright spots or clouding
Micro scratches on the image
plate
i
Replace the image plate.