Troubleshooting
Fault
Probable cause
Solution
X-ray image mirrored in
one corner
Image plate bent during X-ray ex-
posure
i
Do not bend the image plate.
Shadow on the X-ray im-
age
Image plate removed from the
light protection cover before
scanning
i
Do not handle image plates
without a light protection cover.
i
Store the image plate in a light
protection cover.
X-ray image cut off, part
missing
The metal part of the X-ray tube
is in front of the X-ray beam
i
When taking an X-ray, make
sure there are no metal parts
between the X-ray tube and the
patient.
i
Check X-ray tube.
Faulty edge masking in imaging
software
i
Deactivate edge masking.