Troubleshooting
Fault
Probable cause
Solution
Top or bottom bulge in the
X-ray image
Image plate fed in off-centre and
at an angle
i
Check the error code on the
touch screen.
i
Insert the image plate centrally
and straight.
X-ray image is mirror-
inverted
Image plate exposed on the
wrong side.
i
Insert the image plate correctly
in the light protection cover.
i
Position the image plate cor-
rectly.
Round shadow on the X-
ray image
Plus ID image plate (with marker)
exposed on the wrong side
i
When taking an X-ray, make
sure that the active side faces
towards the X-ray tube.
Ghosting or double expo-
sure on X-ray image
Image plate exposed twice
i
Only expose the image plate
once.
Image plate not sufficiently erased
i
Check the erasure unit is work-
ing correctly.
i
Inform a service technician if the
problem persists.
EN