9.5
Fluid container
The fluid container can hold up to approx. 600 ml
of water or water-based fluids.
❯
The fluid container should be emptied and
cleaned every evening and descaled as
required, "12.3 Cleaning the fluid container".
The fluid container is not suitable for ther-
mal disinfection or sterilisation.
Water quality
The water quality must meet the general require-
ments for water supply systems in a dental sur-
gery and all applicable standards.
Watery active ingredients
In order to widen the range of possible treat-
ments, additional active ingredients can be
added to the water.
Possible active ingredients include e.g. chlorhexi-
dine digluconate solutions up to a total concen-
tration of 0.2 %; other solutions may damage the
system and can lead to unsuccessful treatment.
❯
After using a water-based active ingredient the
unit should be thoroughly rinsed using warm
water.
9.6
Cleaning components
Vector/RinsEndo Disinfection
Ready-to-use aldehyde-free solutions for fast
action disinfection and cleaning of Vector hose
systems. The undiluted solution is poured into
the fluid container. These measure must be car-
ried out before breaks in treatment of more than
24 hours, "12.2 Activating the cleaning process
of the device".
Vector cleaner
Ready-to-use solution for the removal of acid-
soluble deposits in the hose system and in the
handpieces of the Vector system. Special cleaner
with intensive cleaning effect and very good
material compatibility.
9.7
Vector toolcard
Instrument wear
The instruments are subject to wear which will
depend on their material,
the surfaces being treated and length of time in
use.
All instruments must be regularly checked with
the Vector toolcard in order to check the degree
of wear, "15.2 Checking instrument wear".
Usage
9000-615-32/02 2012V003
21
EN