RINSE (rinsing/disinfecting):
As a result of the rinsing with water the
special cleaner
Vector cleaner
is removed
from the system. Any residue of the
cleaning agent could cause irritation in the
patient.
❯
Top up the fluid container to the upper mark-
ing with water.
❯
Touch the RINSE (rinsing/disinfecting) button
for at least 2 seconds.
LED RINSE flashes.
RINSE
The device is rinsed with water for 30 seconds.
The process ends automatically.
After the end of the complete cleaning process
(CLEAN + RINSE) the CLEAN LED goes out
again and an acoustic signal sounds (3x high-
pitched audible signal).
If the cleaning process is not performed in full or
it is stopped prematurely, then the blue CLEAN
LED will come on every time the device is
switched on.
12.3 Cleaning the fluid container
The fluid container should regularly cleaned and
descaled.
How soon descaling is required depends primar-
ily on the hardness of the water used. However,
at the latest descaling must be performed when
first signs of limescale become apparent.
Cleaning:
❯
Fill up the fluid container to the upper mark-
ing with cleaning solution.
❯
Allow the cleaning agent to soak as directed
in the manufacturer's information.
❯
Completely empty the fluid container.
❯
Thoroughly rinse the fluid container with
water and dry.
Descaling:
❯
Fill up the fluid container to the upper mark-
ing with e.g. 10% citric acid solution.
❯
Allow the descaler to soak in; follow the
manufacturer's instructions if necessary.
❯
Completely empty the fluid container.
❯
Thoroughly rinse the fluid container with
water and dry.