Check the Scaler handpiece light conductor for
its light transmission, replace if necessary.
13.5 Manual cleaning, intermedi-
ate rinsing, disinfection, final
rinsing, drying in the cleaning
and disinfection bath
A disinfectant or combined cleaning and disinfec-
tant agent is required for manual disinfection. It
must have the following properties:
– certified, possibly virucidal efficacy (DVV/RKI,
VAH or European Standards)
For further information, see: "General informa-
tion".
Cleaning
❯
Place the removable parts of the handpiece
(Scaler cover, light conductor), instrument hold-
ers of the tool kits (without instruments) and
the disassembled handpiece in the cleaning
bath for the required reaction time, so that all
parts are covered.
❯
Clean all accessible external surfaces and inter-
nal surfaces below the surface for 5 minutes
with a hygienic cleaning brush until all visible
soiling has been removed.
❯
Place instruments in small parts baskets into
the cleaning bath.
❯
Rinse handpiece through at least 3 times using
a 20-ml disposable syringe.
20 ml
❯
Screw Scaler instruments onto the rinsing
adapter one after another and rinse through
each of the instruments at least 3 times using a
20-ml disposable syringe.
20 ml
❯
Remove all rinse adapters.
❯
Comply with the cleaning agents' action times
specified by the manufacturer.
Usage
9000-615-32/02 2012V003
35