D-FL 200 / Rev. 6
21
______________________________________________________________________________________________________
Industrie Elektronik GmbH & Co KG
default 332 m/s (corr. to 1089 ft/s)
R32
calibration time
calibration interval (0-720h), disabled=0,
default 24h
R33
normal/reverse
inverse mode for the velocity measurement (0=normal
direction, 1= inverse), default 0
8.1. Key Functions
The keys have the following functions:
Key
For switching to the “
Input” (storing) mode.
The LED on the "STO"-key is illuminated when in this mode.
Key
For switching to the “
Display” mode.
The LED on the "MOD"-key is illuminated.
Key
For counting up when in the “
Input” mode, and for switching the display when
in the “
Display” mode. The LED in the “MOD” key is illuminated.
Key
For counting down when in the “
Input” mode, and for switching back the
display when in the “
Display” mode. The LED in the “MOD”-key is illuminated.
8.2. Storing / Entering Parameters by using the Keys
•
Pressing the “MOD”-key switches from the 'Measuring'-mode to the 'Display'-mode. The “MOD”
LED is illuminated.
•
The “+“-key causes that the parameters will be displayed one after the other, according to the
second table below
Fig.19. The corresponding parameter appears in the display.
•
Pressing the “STO”-key switches the system into the “Input” mode. The “STO” LED is illuminated.
•
The “+“ and “-“ can be used to change the value in the display. The longer these keys are
pressed, the more quickly the value is changed.
•
If the “STO”-key is pressed again, the new value is stored and the system switches back to the
“Display” mode. The “STO” LED is extinguished.
•
By pressing the “MOD”-key the system switches back from the “Display” mode to the “Measuring”
mode. The “MOD” LED is extinguished, and the current measured value is shown on the display.
If the system is switched back from the “Input” mode to the “Measuring” mode using the “MOD”-
key before the new values are stored, they will not be saved.
8.3. Storing / Entering Parameters via the RS232 Interface
Parameters may also be entered using the serial interface of a PC or Notebook. The D_FL_COM
parameter program can run on Windows 3.1x and Windows 95/98. The detailed description how to use
it you find in section
10. PC-Parameterisation of the D-FL 200COM page 25.