SORVALL
®
Centrifuges
Operation
3-7
To prevent rotor failure, the total contents of any compartment, including specimen,
tubes, cover, and adapters (if used), must not exceed the figure given on page 1-2 unless
rotor speed is reduced proportionately.
Strict adherence to the maximum allowable compartment mass or reduced speed is
required to prevent rotor failure.
Observe WARNING on the Safety Information Page
in front of this manual.
The rotor speed is reduced in proportion to the square of the ratio for the maximum
allowable compartment mass to the actual compartment mass (including specimen,
tubes, covers, and adapters). If the compartment mass is more than that specified for the
rotor, the reduced speed can be determined by using the formula given below.
Reduced Rotor Speed = Maximum Rotor Speed x
3-7/3-8
√
Maximum Compartment Mass
Actual Compartment Mass