background image

Drucker Diagnostics – Customer Service: +1-814-692-7661 – 

[email protected]

 | Page 1 

M

ODEL 

D

ESCRIPTION

 

The DASH Coag centrifuge is engineered to reduce turnaround time (TAT) and simplify coag processing in the STAT 
laboratory. Produce error-free samples in as little as 3 minutes with the DASH Coag’s simple set-and-lock controls, 
preset cycle settings, and an LED lid lighting indicator system. 

 

F

EATURES

 

o

 

Simple 2-Button interface 

o

 

Three (3) easily selectable pre-set cycles are conveniently labeled for your lab’s most common applications. Use 

the default cycles or customize them as needed. An LED light indicates the current selected setting.  

o

 

If desired, the control panel can be locked on one preset cycle for error-free reproducibility. 

o

 

Lid lighting indicates the centrifuge’s status (ready, running, done), keeping your TAT down (patent pending). 

o

 

A traditional audible alert indicates the completion of the cycle. 

o

 

Cool–Flow air flow design prevents overheating of samples by maintaining room temperature. 

o

 

Carbon fibers are used to reinforce the tube holders and provide high strength and durability. 

o

 

A clear lid permits safe observation of samples and optical calibration of speed. 

o

 

The lid safety system only allows entry into the centrifuge after the rotor has completely stopped. 

o

 

The lid safety system prevents the centrifuge from operating unless the lid is closed and latched. 

o

 

The high power brushless DC motor provides years of operation with no routine maintenance. 

 

I

NTENDED 

U

SE

 

General purpose laboratory Centrifuge intended for safe and rapid density based separation of fluids, including 
physiologic fluids, in approved specimen receptacles for qualitative or quantitative test procedures. As a general 
purpose laboratory centrifuge, it is designed to also run other approved containers filled with chemicals (non-
flammable, non-explosive, non-volatile, and non-highly reactive only), environmental samples, and other non-
human body samples. This device is intended to be operated by properly trained personnel who have carefully read 
the operating manual and are familiar with the function of the device. 

[Refer to the clinical laboratory method specified by the specimen receptacle manufacturer or established by the 
medical technology for the products applications.]  

 

 

 

Содержание DASH COAG

Страница 1: ...DASH COAG Operator s Manual...

Страница 2: ......

Страница 3: ...ial Setup 2 Quick Start 3 Settings 4 Review Cycle Time and Speed Settings 4 Changing Cycle Time and Speed Settings 4 Operation 5 Balancing Loads 5 Care and Preventative Maintenance 6 Cleaning and Disi...

Страница 4: ......

Страница 5: ......

Страница 6: ......

Страница 7: ...to reinforce the tube holders and provide high strength and durability o A clear lid permits safe observation of samples and optical calibration of speed o The lid safety system only allows entry into...

Страница 8: ...repairs have been performed This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with this operator manual may cause interference to radio c...

Страница 9: ...ng the locking mechanism Entry is only possible when the rotor is stopped Stop Pressing the UNLOCK button during operation will terminate the run and unlock the lid after the rotor has come to astop C...

Страница 10: ...in the current cycle 10 beeps flashes equal 10 minutes of run time Run time starts when the rotor reaches 90 of desired speed and stops when the rotor starts decelerating o Pressing the START button a...

Страница 11: ...will slow to a complete stop and the lid light will flash o The locking mechanism will disengage for 60 seconds allowing entry into the rotor chamber To unlock after more than 60 seconds have elapsed...

Страница 12: ...tingor storing the centrifuge to prevent damage andinjury CLEANING AND DISINFECTION To prolong the life of the centrifuge cleaning and disinfection is recommended every six months or whenever there is...

Страница 13: ...mage o If the tube holders or buckets have any damage even slight safely dispose of them and replace them o If the rotor appears damaged contact Customer Service for further assistance Excessive noise...

Страница 14: ...o change the run preset time The lid does not open o Ensure that the lid knob is turned fully counterclockwise o If the knob cannot be turned counterclockwise turn it fully clockwise press UNLOCK and...

Страница 15: ...0 Watts Centrifuge Motor H P Brushless DC Max G Force 4 400 xg Max Speed 6 600 RPM Min Cycle Time 1 minute Max Cycle Time 30 minutes Use only with approved accessories from the original manufacturer A...

Страница 16: ...ectronic equipment The separate collection and recycling of your waste equipment at the time of disposal will help to conserve natural resources and ensure that it is recycled in a manner that protect...

Отзывы: