Drucker Diagnostics DASH COAG Скачать руководство пользователя страница 9

Drucker Diagnostics – Customer Service: +1-814-692-7661 – 

[email protected]

 | Page 3 

Q

UICK 

S

TART

 

The LED indicator light is on for the cycle currently selected: 

  

2 min 

This setting is factory preset for 2 minutes at 6,600 RPM/4,400 xg 

3 min 

This setting is factory preset for 3 minutes at 6,600 RPM/4,400 xg 

5 min 

This setting is factory preset for 5 minutes at 5,200 RPM/2,700 xg 

 

 

 

 

Start 

Begins running the cycle indicated by the cycle indicator LED light. 

The lid must be closed.

 

 

Unlock 

Allows access into the rotor chamber by disengaging the locking 

mechanism. Entry is only possible when the rotor is stopped. 

 

 

Stop 

Pressing the UNLOCK button during operation will terminate the run 

and unlock the lid after the rotor has come to a stop.

 

 

 

Cycle Selection 

The LED light is on for the cycle currently selected. To change the 

selected cycle, press the UNLOCK button in succession until the 

desired cycle is selected. Two seconds after selection, the button 

reverts to its UNLOCK function. 

 

Lock Cycle Selection 

Select desired cycle. Press and hold the UNLOCK button for 5 

seconds. Two beeps will confirm that cycle selection is locked. 

 

Unlock Cycle Selection 

To re-enable cycle selection, press and hold the UNLOCK button for 

5 seconds. Three beeps will confirm that cycle selection is now 

unlocked. 

 

 

Содержание DASH COAG

Страница 1: ...DASH COAG Operator s Manual...

Страница 2: ......

Страница 3: ...ial Setup 2 Quick Start 3 Settings 4 Review Cycle Time and Speed Settings 4 Changing Cycle Time and Speed Settings 4 Operation 5 Balancing Loads 5 Care and Preventative Maintenance 6 Cleaning and Disi...

Страница 4: ......

Страница 5: ......

Страница 6: ......

Страница 7: ...to reinforce the tube holders and provide high strength and durability o A clear lid permits safe observation of samples and optical calibration of speed o The lid safety system only allows entry into...

Страница 8: ...repairs have been performed This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with this operator manual may cause interference to radio c...

Страница 9: ...ng the locking mechanism Entry is only possible when the rotor is stopped Stop Pressing the UNLOCK button during operation will terminate the run and unlock the lid after the rotor has come to astop C...

Страница 10: ...in the current cycle 10 beeps flashes equal 10 minutes of run time Run time starts when the rotor reaches 90 of desired speed and stops when the rotor starts decelerating o Pressing the START button a...

Страница 11: ...will slow to a complete stop and the lid light will flash o The locking mechanism will disengage for 60 seconds allowing entry into the rotor chamber To unlock after more than 60 seconds have elapsed...

Страница 12: ...tingor storing the centrifuge to prevent damage andinjury CLEANING AND DISINFECTION To prolong the life of the centrifuge cleaning and disinfection is recommended every six months or whenever there is...

Страница 13: ...mage o If the tube holders or buckets have any damage even slight safely dispose of them and replace them o If the rotor appears damaged contact Customer Service for further assistance Excessive noise...

Страница 14: ...o change the run preset time The lid does not open o Ensure that the lid knob is turned fully counterclockwise o If the knob cannot be turned counterclockwise turn it fully clockwise press UNLOCK and...

Страница 15: ...0 Watts Centrifuge Motor H P Brushless DC Max G Force 4 400 xg Max Speed 6 600 RPM Min Cycle Time 1 minute Max Cycle Time 30 minutes Use only with approved accessories from the original manufacturer A...

Страница 16: ...ectronic equipment The separate collection and recycling of your waste equipment at the time of disposal will help to conserve natural resources and ensure that it is recycled in a manner that protect...

Отзывы: