Drucker Diagnostics – Customer Service: +1-814-692-7761 –
S
ETTINGS
Setting 1
Setting 2
Setting 3
RPM
6,600
6,600
5,200
Time
2 min
3 min
5 min
G-Force
4,400 xg
4,400 xg
2,700 xg
R
EVIEW
C
YCLE
T
IME AND
S
PEED
S
ETTINGS
Your settings may not be standard. Factory programmed cycles are shown on the rear of the centrifuge, on the
Factory Set Cycles label. To review current settings, follow this procedure:
o
The lid must be open to review the selected cycle time and speed.
o
Press and hold the START button until you hear a beep.
o
Release the START button. The centrifuge will beep and the LED light will flash once for each minute of run time
in the current cycle. 10 beeps / flashes equal 10 minutes of run time. Run time starts when the rotor reaches 90%
of desired speed and stops when the rotor starts decelerating.
o
Pressing the START button again will cause the unit to beep and the LED light to flash once for each 100
revolutions per minute (RPM) in the current cycle. 10 beeps / flashes equal 10 x 100 or 1,000 RPM.
o
The centrifuge will automatically revert to normal mode at the end.
C
HANGING
C
YCLE
T
IME AND
S
PEED
S
ETTINGS
o
The lid must be open to change the selected cycle time and speed.
o
Select the cycle you wish to change.
o
Press and hold the START and UNLOCK buttons together until the LED light flashes.
o
Press the START button for each minute of run time.
o
Move to speed setting mode by
pressing the UNLOCK button.
o
Press the START button once for each 100 rpm.
o
Press the UNLOCK button to exit setting mode.