DRUCK
TEMPERATUR
&
Leitenberger GmbH •
Bahnhofstr. 33 • D-72138 Kirchentellinsfurt •
Tel. +49 (0) 7121-90920-0 • Fax +49 (0) 7121-90920-99 •
• www.druck-temperatur.de
GERMANY
DT
www.
-Cal.net
LR
MANUAL
LPC 300 Druckkalibrator
300 Pressure Calibrator
LPC
90
Operang mode:
-
SWITCH TEST
In the
-
mode, along with the
-S reference pressure sensor data (see
SWITCH TEST
LPC
MEASURING mode), the status and switching point of the pressure switch are also displayed.
a) Current measured value of the
reference pressure sensor
LPC-S
b) Current switch posion/status of the test item (pressure switch)
c) Opening switching point
d) Closing switching point
e) Hysteresis/separaon between opening and closing of the switch (test item)
f) Pressure unit (se able via menu)
SWITCH TEST
-
0.000 10.000
4.200
bar
Status:
4.900
5.000
bar
bar
Hys:
0.100
bar
Reference
a)
b)
c)
d)
e)
f)