DRUCK
TEMPERATUR
&
Leitenberger GmbH •
Bahnhofstr. 33 • D-72138 Kirchentellinsfurt •
Tel. +49 (0) 7121-90920-0 • Fax +49 (0) 7121-90920-99 •
• www.druck-temperatur.de
GERMANY
DT
www.
-Cal.net
LR
MANUAL
LPC 300 Druckkalibrator
300 Pressure Calibrator
LPC
105
In order to switch the calibrator into
-
mode, follow the instrucons on the previous
page.
SWITCH TEST
The
-
funcon is
suitable for electronic switches (e.g.
- or
SWITCH TEST
NOT
PNP
switches), but is only for mechanical, potenal-free switches.
NPN
The following is a more detailed explanaon of point „2. Preparaon for
-
“
SWITCH TEST
a) Voltage supply for test item (on/off). If no external supply is required for the test item, „
“
OFF
should be selected to conserve energy.
b) Unit and resoluon (submenu)
The calculated measured values of the two switching points and the hysteresis
can be reset by pressing the „0" (zero) key on the numeric keypad.
Customer-specific unit; with respect to bar
(input via numeric
)
ENTER
Display resoluon in operaing mode
(back with
)
Select and confirm (standard units) via
BACK
Invers: aktuelle Cursor-Posion. Wechsel mit , Auswahl bestägen mit
xxx
vw tu
Parameter-Auswahl aus Liste bzw. Menü mit , Auswahl bestägen mit
0.000
Parameter-Eingabe mit numerischem Eingabeblock, bestägen mit
CLEAR
Löschen einer Eingabe mit
K
urzin
fo