3. Specificaons
LR-
Cal
300
LPC
Type of pressure
gauge pressure, vacuum, compound ranges, absolute
Pressure units
15 standard pressure units + 1 freely programmable unit
Acve temperature compensaon
0…50°C
Permissible ambient temperature
0…50°C
Calibraon
Traceable factory cerficate of calibraon (Oponal:
/
kkS)
DKD DA
Number of supported
LPC-S
sensors up to 10
LPC-S
reference pressure sensors
Accuracy of measuring chain
Ranges up to 1000 bar:
±0.025% FS
(at 23°C, vercal mounng)
Ranges >1000 bar:
±0,1% FS
(at 23°C, vercal mounng)
Operang modes
MEASURE CALIBRATE SWITCH TEST
/
/
-
Display
Large
colour screen for the display of reference and test
TFT
signals, and addional informaon
Display resoluon
up to 6 digits, selectable
Measuring rate (pressure)
2 values/sec
Funcons
CALIBRATE SWITCH TEST
,
-
, Min/Max-memory, Tare, Min-/Max-
Alarm(visual), Filter (running average), Zero-Point adjustment,
powersafe funcon
CALIBRATION funcon
Memory capacity up to 16 test items
Test points per test item
up to 32 comparison points
SWITCH TEST
-
funcon
Determinaon of the switch point and automac calculaon
of the hysteresis
Measuring input VOLTAGE
Measuring range
0…1, 0…2, 0…5, 0…10 V
Resoluon
0.1 mV
Accuracy
1 mV
Measuring input CURRENT
Measuring range
0…20, 4…20 mA
Resoluon
0.1 μA
Accuracy
5 μA
Loop supply voltage
24 V (Load max. 50 mA, min. 20 mA), can be acvated via menu
Interfaces
RS
USB
232 and
Current supply
Internal Lithium-Ion rechargeable ba ery (charging me <6 h)
Ba ery operaon
approx. 20 h
Persmissible relave humidity
0…85% r.h. (non-condensing, at 50°C)
Permissible storage temperature
-20...+70°C
Case
Impact-resistant
plasc, membrane keypad,
ABS
transparent screen
Ingress protecon
IP 54 (only with oponal caps closed)
Weight
approx. 850 g (without sensor)
CE conformity
EMC
EC EN
direcve 2004/108
,
61326 emission (group 1, class B)
and immunity (portable equipment)