Drive DeVilbiss Healthcare Theia Скачать руководство пользователя страница 2

2

MUTE

STATIC

POWER FAILURE

LOW PRESSURE

WEIGHT RANGE

1

MIN

MAX

MID

4

8

• 

Mattress only - For profiling beds, it is 

essential that straps are secured around the 

movable sections of the bed frame - damage 

will be incurred when profiled if secured to 

fixed parts of the frame

 (FIG 1).

• 

Cushion only:  It is the responsibility of the 

carer to ensure the chair is suitable for 

product compatibility and the safety of the 

patient. Position the cushion with the pipes at 

the rear of the chair

• 

To avoid any risk of damage to the mattress/

cushion ensure there are no sharp objects 

which may come into contact with it.

• 

Position the control unit by hanging the 

hooks over the foot board. If there is no foot 

board or the cushion is in use place the unit 

on the floor with the front facing upwards. 

Ensure the rear of the unit is not obstructed 

by carpet, rugs etc. It is advisable to place the 

unit on a firm surface 

(FIG 2).

• 

The mattress /cushion will start to inflate. 

Inflation can take up to 40mins. Once inflated, 

ensure the straps attaching the mattress 

to the bed frame are secure and hold the 

mattress in place. Secure sheets loosely 

enough to ensure they do not interfere with 

cell alternation.

CPR - mattress only

• 

Rapid deflation of the mattress may be required 

for emergency treatment or system deflation. 

The CPR pull tag is located at the foot end of 

the mattress.

• 

Pull the CPR tag to disengage from the 

mattress, once done the entire system will 

rapidly deflate.

• 

To re-inflate, re attach the CPR tag 

(FIG 4)

.

• 

Wait for the mattress system to reach optimal 

pressure prior to a return to normal use.

QUICK REFERENCE GUIDE P1 of 2

FIG 1.

FIG 2.

FIG 3.

FIG 4.

• 

Connect the mattress/cushion to the control 

unit 

(FIG 5) 

ensuring the feed pipes do not 

kink or become trapped between bed/chair 

parts.  

• 

Switch pump on by the ON/OFF switch 

(FIG 3)

.

• 

Mattresses - ensure that the CPR tag is 

fully engaged within the mattress and all 

connectors are lined up correctly 

(FIG 4).

TO BE USED IN CONJUNCTION WITH THE FULL INSTRUCTIONS FOR USE FOR THIS 

PRODUCT. THIS QUICK GUIDE DOES NOT REPLACE THE FULL DOCUMENT.

THEIA

 

Pump with mattress options and cushion

Overlay Mattress

Air on Foam Mattress

Dynamic Cushion

The mattress is intended to support a 

single patient who is up to 180kg in weight 

and 185cm in height. For those patients 

of a very low weight, typically less than 

40kg and a physical size less than 146cm, 

clinical judgment is to be used to determine 

suitability. A lower (or upper) age limit is not 

defined as it depends on the physical size of 

the patient in relation to the proportions of 

the mattress and bed frame.

The mattress is intended to support a 

single patient who is up to 180kg in weight 

and 185cm in height. For those patients 

of a very low weight, typically less than 

40kg and a physical size less than 146cm, 

clinical judgment is to be used to determine 

suitability. A lower (or upper) age limit is not 

defined as it depends on the physical size of 

the patient in relation to the proportions of 

the mattress and bed frame.

The cushion is intended to support a single 

patient who is up to 120kg in weight. A 

lower (or upper) age limit is not defined 

as it depend on the physical size of the 

patient in relation to the proportions of 

the cushion and the platform. Clinical 

judgement is to be used to determine 

patient suitability.

Содержание DeVilbiss Healthcare Theia

Страница 1: ...Dynamic System Instructions for use...

Страница 2: ...o normal use QUICK REFERENCE GUIDE P1 of 2 FIG 1 FIG 2 FIG 3 FIG 4 Connect the mattress cushion to the control unit FIG 5 ensuring the feed pipes do not kink or become trapped between bed chair parts...

Страница 3: ...and cushion MUTE STATIC POWER FAILURE LOW PRESSURE WEIGHT RANGE 1 MIN MAX MID 4 8 No Symbol Description 3 MUTE STATIC POWER FAILURE LOW PRESSURE WEIGHT RANGE 1 MIN MAX MID 4 8 Static Mode Press to set...

Страница 4: ...1 Preparing for Use 8 2 Control Interface 8 3 CPR Function 8 4 Low Pressure Warning 8 5 Power Failure Warning 9 DECONTAMINATION CLEANING 9 1 Control Unit 9 2 Mattress and Cushion 9 3 Alternative Cove...

Страница 5: ...ilbiss Healthcare Ltd service engineers or qualified personnel 1 INTRODUCTION 2 CONTACT INFORMATION Drive DeVilbiss Healthcare Ltd Sidhil Business Park Holmfield Halifax HX2 9TN United Kingdom For Ser...

Страница 6: ...as it depends on the physical size of the patient in relation to the proportions of the mattress and bed frame 3 3 Intended Use The intended use of the support surface is for sleeping resting sitting...

Страница 7: ...tures 3 5 Indications To assist as part of an overall programme of care when active load distribution through mechanical means is required Bed frames used with the system can vary greatly depending on...

Страница 8: ...dema when using alternating mode only Other contraindications may be relevant which are specific to the patient or care environment 4 4 Training If these instructions for use are not deemed sufficient...

Страница 9: ...A hazard could be introduced Occupants and users of this equipment must never smoke in close proximity to the control unit mattress cushion or bedding being used with it Risk of fire Accessories that...

Страница 10: ...could be incurred Do not store whilst inflated damage could be incurred Do not store objects such as side rails on top of the mattress or cushion damage could be incurred Warning Caution 5 2 Transport...

Страница 11: ...e of potential product damage Manufacturer Date of manufacture Lot number Class II electrical device Theuser occupantisprotectedbyatleasttwolayersofinsulationbetweenthecurrent carrying parts e g contr...

Страница 12: ...ts Type BF Applied parts which are electrically isolated from earth and other parts of the medical equipment Complying with specific requirements for protection against electric shock to EN 60601 1 95...

Страница 13: ...electrical cables can create a risk of electrocution and or fire A CE marked extension cable must only be used when it is not possible to reach a wall socket with the equipment mains cable Contact Dr...

Страница 14: ...imum of 2 hours prior to plugging into the mains supply Risk of electrical system damage if operated outside of the recommended temperatures Caution Open all packaging with care Once removed from the...

Страница 15: ...control unit ensuring the air hose does not kink or become trapped between parts of the bed frame Plug the mains cable into a suitable mains supply and switch on the control unit The system will star...

Страница 16: ...ay It s recommended to adjust comfort control to the max when inflating before adjusting the pressure to the patient s needs Before changing the comfort control a clinical judgement is required from f...

Страница 17: ...connections are secure and correctly installed according to the relevant instructions If the pressure is consistently low open the zipper and confirm that all of the hoses are properly connected Chec...

Страница 18: ...of the below washing instructions are not followed the product warranty will be invalidated Do not use solvents neat bleach phenolic based cleaning solutions or abrasive products to clean the casing...

Страница 19: ...ter and a clean cloth and allow to fully dry before use Ensure the internal foam and air cells are orientated correctly to the mattress covers when reassembling 9 3 Alternative Cover Cleaning Instruct...

Страница 20: ...ble and visual indicators work appropriately when plugged in and unplugged from mains supply Check that the battery is still functional and operates in the event of a power loss Check that the mattres...

Страница 21: ...used with your system is not to be disposed of in general municipal waste as it is to be considered as Waste Electrical and Electronic Equipment WEEE Some of the electrical components could be harmfu...

Страница 22: ...Cushion Weight 2kg Maximum Patient Weight 120kg 18 stone No of Cells 6 cells Alternating Therapy AB pattern Cycle Time 12 mins Pressure Range 20 55mmHg Control Unit Dimensions L 270mm x W 155mm x H 10...

Страница 23: ...Mattress Weight 9 7kg Maximum Patient Weight 180kg 28 stone No of Cells 17 cells Alternating Therapy AB pattern Cycle Time 12 mins Pressure Range 20 55mmHg Control Unit Dimensions L 270mm x W 155mm x...

Страница 24: ...Weight 180kg 28 stone No of Cells 17 cells Alternating Therapy AB pattern Cycle Time 12 mins Pressure Range 20 55mmHg Control Unit Dimensions L 270mm x W 155mm x H 100mm Control Unit Weight 1 6kg Cove...

Страница 25: ...ed for domestic purposes Harmonic emissions IEC 61000 3 2 Class A Voltage fluctuations flicker emissions IEC 61000 3 3 Complies Guidance and manufacture s declaration electromagnetic immunity Immunity...

Страница 26: ...For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m can be estimated using the equation applicable to the frequency of the transmitter...

Страница 27: ...urer and d is the recommended separation distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site surveya should be less than the compliance level in ea...

Страница 28: ...ow instructions or g otherwise used in a manner inconsistent with any instructions provided by Drive DeVilbiss Healthcare Ltd The warranty explicitly exempts consumable items Thiswarrantycontainstheen...

Страница 29: ...29...

Страница 30: ...Airflo xiamen Medical Co Ltd 1F 3F 4F No 6 East Haijing Road Haicang Xiamen Fujian China Y Sung Handelsvertretung Duesselihaler Str 24 40211 Duesseldorf Germany Drive DeVilbiss Healthcare Ltd Sidhil...

Отзывы: