9. DECONTAMINATION/CLEANING

Infection control and routine cleaning must be carried out in accordance with 

your local infection control policy or regulatory body. 

Warning

•  Always disconnect the support surface and bed frame from 

the main power supply prior to cleaning.

•  The control unit is rated to IP21 and provides protection 

from condensation only, do not immerse or soak the control 

unit – Risk of electric shock.

•  Regular cleaning and disinfection of the support surface      

will help to prevent the risk of infection to the occupant and/

or carer. 

•  Prior to transferring the system to another user ensure it has 

been cleaned and disinfected using the method as detailed 

below to help prevent the risk of cross infection.

•  Deviations from the specified cleaning and disinfection 

instructions can cause serious hazards, and adversely affect 

the life and efficacy of the system.

•  If any of the below washing instructions are not followed the 

product warranty will be invalidated.

•  Do not use solvents, neat bleach, phenolic based cleaning 

solutions or abrasive products to clean the casing or 

mattress.

Caution

9.1 

Control Unit

•  Check for external damage – If damaged take the control unit out of use.
•  All surfaces to be wiped down with a disposable soft cloth moistened 

with a mild detergent and diluted in warm water (40°C). 

•  The control unit is be cleaned by starting with the cleanest parts of it and 

systematically moving to the dirtiest parts. Extra care should be taken 

around areas where excess dirt or dust may gather. 

•  The cloth should be changed during the cleaning process if it becomes 

soiled. 

•  Wipe down with a clean cloth moistened with clean water to remove 

detergent residue. 

•  If there are blood spillages or bodily fluids present wipe surfaces down 

with 0.1% Chlorine solution (1,000 ppm). 

•  Wipe down with a clean cloth moistened with clean water.
•  Dry off with a paper towel - Always ensure the cleaned surfaces are 

allowed to fully dry before putting back into use.

Содержание DeVilbiss Healthcare Theia

Страница 1: ...Dynamic System Instructions for use...

Страница 2: ...o normal use QUICK REFERENCE GUIDE P1 of 2 FIG 1 FIG 2 FIG 3 FIG 4 Connect the mattress cushion to the control unit FIG 5 ensuring the feed pipes do not kink or become trapped between bed chair parts...

Страница 3: ...and cushion MUTE STATIC POWER FAILURE LOW PRESSURE WEIGHT RANGE 1 MIN MAX MID 4 8 No Symbol Description 3 MUTE STATIC POWER FAILURE LOW PRESSURE WEIGHT RANGE 1 MIN MAX MID 4 8 Static Mode Press to set...

Страница 4: ...1 Preparing for Use 8 2 Control Interface 8 3 CPR Function 8 4 Low Pressure Warning 8 5 Power Failure Warning 9 DECONTAMINATION CLEANING 9 1 Control Unit 9 2 Mattress and Cushion 9 3 Alternative Cove...

Страница 5: ...ilbiss Healthcare Ltd service engineers or qualified personnel 1 INTRODUCTION 2 CONTACT INFORMATION Drive DeVilbiss Healthcare Ltd Sidhil Business Park Holmfield Halifax HX2 9TN United Kingdom For Ser...

Страница 6: ...as it depends on the physical size of the patient in relation to the proportions of the mattress and bed frame 3 3 Intended Use The intended use of the support surface is for sleeping resting sitting...

Страница 7: ...tures 3 5 Indications To assist as part of an overall programme of care when active load distribution through mechanical means is required Bed frames used with the system can vary greatly depending on...

Страница 8: ...dema when using alternating mode only Other contraindications may be relevant which are specific to the patient or care environment 4 4 Training If these instructions for use are not deemed sufficient...

Страница 9: ...A hazard could be introduced Occupants and users of this equipment must never smoke in close proximity to the control unit mattress cushion or bedding being used with it Risk of fire Accessories that...

Страница 10: ...could be incurred Do not store whilst inflated damage could be incurred Do not store objects such as side rails on top of the mattress or cushion damage could be incurred Warning Caution 5 2 Transport...

Страница 11: ...e of potential product damage Manufacturer Date of manufacture Lot number Class II electrical device Theuser occupantisprotectedbyatleasttwolayersofinsulationbetweenthecurrent carrying parts e g contr...

Страница 12: ...ts Type BF Applied parts which are electrically isolated from earth and other parts of the medical equipment Complying with specific requirements for protection against electric shock to EN 60601 1 95...

Страница 13: ...electrical cables can create a risk of electrocution and or fire A CE marked extension cable must only be used when it is not possible to reach a wall socket with the equipment mains cable Contact Dr...

Страница 14: ...imum of 2 hours prior to plugging into the mains supply Risk of electrical system damage if operated outside of the recommended temperatures Caution Open all packaging with care Once removed from the...

Страница 15: ...control unit ensuring the air hose does not kink or become trapped between parts of the bed frame Plug the mains cable into a suitable mains supply and switch on the control unit The system will star...

Страница 16: ...ay It s recommended to adjust comfort control to the max when inflating before adjusting the pressure to the patient s needs Before changing the comfort control a clinical judgement is required from f...

Страница 17: ...connections are secure and correctly installed according to the relevant instructions If the pressure is consistently low open the zipper and confirm that all of the hoses are properly connected Chec...

Страница 18: ...of the below washing instructions are not followed the product warranty will be invalidated Do not use solvents neat bleach phenolic based cleaning solutions or abrasive products to clean the casing...

Страница 19: ...ter and a clean cloth and allow to fully dry before use Ensure the internal foam and air cells are orientated correctly to the mattress covers when reassembling 9 3 Alternative Cover Cleaning Instruct...

Страница 20: ...ble and visual indicators work appropriately when plugged in and unplugged from mains supply Check that the battery is still functional and operates in the event of a power loss Check that the mattres...

Страница 21: ...used with your system is not to be disposed of in general municipal waste as it is to be considered as Waste Electrical and Electronic Equipment WEEE Some of the electrical components could be harmfu...

Страница 22: ...Cushion Weight 2kg Maximum Patient Weight 120kg 18 stone No of Cells 6 cells Alternating Therapy AB pattern Cycle Time 12 mins Pressure Range 20 55mmHg Control Unit Dimensions L 270mm x W 155mm x H 10...

Страница 23: ...Mattress Weight 9 7kg Maximum Patient Weight 180kg 28 stone No of Cells 17 cells Alternating Therapy AB pattern Cycle Time 12 mins Pressure Range 20 55mmHg Control Unit Dimensions L 270mm x W 155mm x...

Страница 24: ...Weight 180kg 28 stone No of Cells 17 cells Alternating Therapy AB pattern Cycle Time 12 mins Pressure Range 20 55mmHg Control Unit Dimensions L 270mm x W 155mm x H 100mm Control Unit Weight 1 6kg Cove...

Страница 25: ...ed for domestic purposes Harmonic emissions IEC 61000 3 2 Class A Voltage fluctuations flicker emissions IEC 61000 3 3 Complies Guidance and manufacture s declaration electromagnetic immunity Immunity...

Страница 26: ...For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m can be estimated using the equation applicable to the frequency of the transmitter...

Страница 27: ...urer and d is the recommended separation distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site surveya should be less than the compliance level in ea...

Страница 28: ...ow instructions or g otherwise used in a manner inconsistent with any instructions provided by Drive DeVilbiss Healthcare Ltd The warranty explicitly exempts consumable items Thiswarrantycontainstheen...

Страница 29: ...29...

Страница 30: ...Airflo xiamen Medical Co Ltd 1F 3F 4F No 6 East Haijing Road Haicang Xiamen Fujian China Y Sung Handelsvertretung Duesselihaler Str 24 40211 Duesseldorf Germany Drive DeVilbiss Healthcare Ltd Sidhil...

Отзывы: