5. TRANSPORT AND STORAGE

5.1 Storage

•  Detach the control unit from the support surface.
•  Release CPR or rotate the CPR dial until it is open.
•  Lay the mattress or cushion out flat and position upside down. 
•  Ensure there is no air trapped in the cells.
•  Position the control unit on the mattress or cushion.
•  The product should be rolled from the head end towards the foot end 

(ensuring the control unit is fully covered).

•  Place into storage bag/holdall to protect from dirt, debris, fluids etc.

• 

To prevent the risk of cross infection, when removing the system from 

an end user’s residence ensure that all activities in relation to the 

system are carried out using disposable gloves and that they are then 

discarded appropriately, unless it can be verified that the cushion, 

mattress and control unit have been suitably cleaned and disinfected 

prior to collection.

• 

On the return of the system from an end users residence, prior to 

putting into storage ensure it has been cleaned and disinfected in line 

with the local infection control policy and/or as defined in section 10 

of these instructions for use.

• 

Do not crease or stack mattresses, cushion and/or control units - 

damage could be incurred.

• 

Do not store whilst inflated - damage could be incurred.

• 

Do not store objects such as side rails on top of the mattress or 

cushion - damage could be incurred.

Warning

Caution

5.2 Transportation

Where possible, it is recommended the transport of mattresses should be 

carried out on a flat based trolley or mattress trolley. Do not drag or pull the 

mattress by its cover or foam core. Please follow local moving and handling 

policies and guidelines when handling a mattress. It is recommended that two 

people manoeuvre the mattress.

You can achieve 8 hours transport time by 

carrying out the following procedure:
• Disconnect the umbilical cord from the 

control unit by squeezing the two tabs and 

pulling away (FIG 5).
• Seal using the cap marked ‘Transport’ 

(FIG 6).
• Switch off the control unit.

QUICK REFERENCE GUIDE P2 of 2

TO BE USED IN CONJUNCTION WITH THE FULL INSTRUCTIONS FOR USE FOR THIS 

PRODUCT. THIS QUICK GUIDE DOES NOT REPLACE THE FULL DOCUMENT.

No. Symbol

Description

1

LOW PRESSURE

NORMAL PRESSURE

MIN

MAX

W

EI

GH

RA

NG

E

MID

MUTE

STATIC

Weight Range Control:

 Turn dial 

clockwise to increase mattress 

pressure and anti-clockwise to 

decrease mattress pressure. It 

is recommended to adjust the 

control to the max when first 

inflated. Before changing the 

control a clinical judgement is 

required.

2

LOW PRESSURE

NORMAL PRESSURE

MIN

MAX

W

EI

GH

RA

NG

E

MID

MUTE

STATIC

Static Mode: 

Press to set the 

air within the mattress to stay 

static, green light will indicate 

static mode is selected. The 

system will not revert back to 

alternating mode. The user must 

press alternate button to return 

to alternating mode.      

Drive DeVilbiss Sidhil Ltd     

|

    APOLLO II   

|

   Quick Reference guide   

|

    REV 5

EROS

 

Pump with mattress options and cushion

LOW PRESSURE

NORMAL PRESSURE

MIN

MAX

W

EI

GH

RA

NG

E

MID

MUTE

STATIC

3

2

1

No. Symbol

Description

3

LOW PRESSURE

NORMAL PRESSURE

MIN

MAX

W

EI

GH

RA

NG

E

MID

MUTE

STATIC

Alarm Mute: 

The audible alarm 

will sound when the pressure 

within the system is low or a 

power failure is detected. To 

mute, press the mute button and 

to reactivate the alarm, re-press 

the mute button.

LOW PRESSURE

NORMAL PRESSURE

MIN

MAX

W

EI

GH

RA

NG

E

MID

MUTE

STATIC

Low Pressure :

 When abnormal 

low pressure occurs, the low 

pressure indicator will illuminate 

and the alarm will be activated 

to alert of a low pressure 

condition.

Transport Mode

You can achieve 8 hours transport time by 

carrying out the following procedure:

• 

Disconnect the umbilical cord from the power 

unit by squeezing the two tabs and pulling 

away 

(FIG 5).

• 

Seal using the cap marked ‘Transport’ 

(FIG 6).

• 

Switch off the control unit.

1

1

2

FIG 5.

FIG 6.

Содержание DeVilbiss Healthcare Theia

Страница 1: ...Dynamic System Instructions for use...

Страница 2: ...o normal use QUICK REFERENCE GUIDE P1 of 2 FIG 1 FIG 2 FIG 3 FIG 4 Connect the mattress cushion to the control unit FIG 5 ensuring the feed pipes do not kink or become trapped between bed chair parts...

Страница 3: ...and cushion MUTE STATIC POWER FAILURE LOW PRESSURE WEIGHT RANGE 1 MIN MAX MID 4 8 No Symbol Description 3 MUTE STATIC POWER FAILURE LOW PRESSURE WEIGHT RANGE 1 MIN MAX MID 4 8 Static Mode Press to set...

Страница 4: ...1 Preparing for Use 8 2 Control Interface 8 3 CPR Function 8 4 Low Pressure Warning 8 5 Power Failure Warning 9 DECONTAMINATION CLEANING 9 1 Control Unit 9 2 Mattress and Cushion 9 3 Alternative Cove...

Страница 5: ...ilbiss Healthcare Ltd service engineers or qualified personnel 1 INTRODUCTION 2 CONTACT INFORMATION Drive DeVilbiss Healthcare Ltd Sidhil Business Park Holmfield Halifax HX2 9TN United Kingdom For Ser...

Страница 6: ...as it depends on the physical size of the patient in relation to the proportions of the mattress and bed frame 3 3 Intended Use The intended use of the support surface is for sleeping resting sitting...

Страница 7: ...tures 3 5 Indications To assist as part of an overall programme of care when active load distribution through mechanical means is required Bed frames used with the system can vary greatly depending on...

Страница 8: ...dema when using alternating mode only Other contraindications may be relevant which are specific to the patient or care environment 4 4 Training If these instructions for use are not deemed sufficient...

Страница 9: ...A hazard could be introduced Occupants and users of this equipment must never smoke in close proximity to the control unit mattress cushion or bedding being used with it Risk of fire Accessories that...

Страница 10: ...could be incurred Do not store whilst inflated damage could be incurred Do not store objects such as side rails on top of the mattress or cushion damage could be incurred Warning Caution 5 2 Transport...

Страница 11: ...e of potential product damage Manufacturer Date of manufacture Lot number Class II electrical device Theuser occupantisprotectedbyatleasttwolayersofinsulationbetweenthecurrent carrying parts e g contr...

Страница 12: ...ts Type BF Applied parts which are electrically isolated from earth and other parts of the medical equipment Complying with specific requirements for protection against electric shock to EN 60601 1 95...

Страница 13: ...electrical cables can create a risk of electrocution and or fire A CE marked extension cable must only be used when it is not possible to reach a wall socket with the equipment mains cable Contact Dr...

Страница 14: ...imum of 2 hours prior to plugging into the mains supply Risk of electrical system damage if operated outside of the recommended temperatures Caution Open all packaging with care Once removed from the...

Страница 15: ...control unit ensuring the air hose does not kink or become trapped between parts of the bed frame Plug the mains cable into a suitable mains supply and switch on the control unit The system will star...

Страница 16: ...ay It s recommended to adjust comfort control to the max when inflating before adjusting the pressure to the patient s needs Before changing the comfort control a clinical judgement is required from f...

Страница 17: ...connections are secure and correctly installed according to the relevant instructions If the pressure is consistently low open the zipper and confirm that all of the hoses are properly connected Chec...

Страница 18: ...of the below washing instructions are not followed the product warranty will be invalidated Do not use solvents neat bleach phenolic based cleaning solutions or abrasive products to clean the casing...

Страница 19: ...ter and a clean cloth and allow to fully dry before use Ensure the internal foam and air cells are orientated correctly to the mattress covers when reassembling 9 3 Alternative Cover Cleaning Instruct...

Страница 20: ...ble and visual indicators work appropriately when plugged in and unplugged from mains supply Check that the battery is still functional and operates in the event of a power loss Check that the mattres...

Страница 21: ...used with your system is not to be disposed of in general municipal waste as it is to be considered as Waste Electrical and Electronic Equipment WEEE Some of the electrical components could be harmfu...

Страница 22: ...Cushion Weight 2kg Maximum Patient Weight 120kg 18 stone No of Cells 6 cells Alternating Therapy AB pattern Cycle Time 12 mins Pressure Range 20 55mmHg Control Unit Dimensions L 270mm x W 155mm x H 10...

Страница 23: ...Mattress Weight 9 7kg Maximum Patient Weight 180kg 28 stone No of Cells 17 cells Alternating Therapy AB pattern Cycle Time 12 mins Pressure Range 20 55mmHg Control Unit Dimensions L 270mm x W 155mm x...

Страница 24: ...Weight 180kg 28 stone No of Cells 17 cells Alternating Therapy AB pattern Cycle Time 12 mins Pressure Range 20 55mmHg Control Unit Dimensions L 270mm x W 155mm x H 100mm Control Unit Weight 1 6kg Cove...

Страница 25: ...ed for domestic purposes Harmonic emissions IEC 61000 3 2 Class A Voltage fluctuations flicker emissions IEC 61000 3 3 Complies Guidance and manufacture s declaration electromagnetic immunity Immunity...

Страница 26: ...For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m can be estimated using the equation applicable to the frequency of the transmitter...

Страница 27: ...urer and d is the recommended separation distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site surveya should be less than the compliance level in ea...

Страница 28: ...ow instructions or g otherwise used in a manner inconsistent with any instructions provided by Drive DeVilbiss Healthcare Ltd The warranty explicitly exempts consumable items Thiswarrantycontainstheen...

Страница 29: ...29...

Страница 30: ...Airflo xiamen Medical Co Ltd 1F 3F 4F No 6 East Haijing Road Haicang Xiamen Fujian China Y Sung Handelsvertretung Duesselihaler Str 24 40211 Duesseldorf Germany Drive DeVilbiss Healthcare Ltd Sidhil...

Отзывы: