EN - 96
SE-7314-2
Disinfection
NOTE
– The disinfection process can only be completed by the manufacturer or by an appropriately trained individual.
Recommended
disinfection interval
Number of
disinfection cycles
Compatible disinfection method
Enclosure, power cord
Between patients
20
1:5 chlorine bleach (5.25%) and water solution,
Mikrobac forte, Terralin Protect
Collection container, filter, tubing and carrying case
Do not clean, replace between
patients
N/A
N/A
PROVIDER’S NOTES
No routine calibration or service is required provided the device is used in accordance with the manufacturer’s directions. In case of a change of patient, the device must be
reconditioned to protect the user. Reconditioning must only be carried out by the manufacturer or service provider. Between patients:
1. Visually inspect unit for any damage, missing parts etc.
2. Ensure that unit and accessories are clean.
3. Use an independent vacuum gauge to verify the unit provides the proper vacuum level as stated in Specifications.
4. Discard and replace collection container, filter, tubing, and carrying case between patients.
5. Wipe the surface using a clean cloth dampened with disinfectant.
TROUBLESHOOTING
DANGER
Electric shock hazard. Do not attempt to open or remove cabinet, there are no user-serviceable internal components. If service is required, return unit to a
qualified DeVilbiss Healthcare provider or an authorized service center. Opening or tampering with the unit will void warranty.
NOTE
– Your DeVilbiss Suction Unit contains no user-serviceable parts. If you believe your unit is not working properly, BEFORE YOU RETURN IT TO your DeVilbiss
Healthcare PROVIDER WHERE YOU PURCHASED IT, please take a few moments to check for these possible causes:
PROBLEM
ACTION
Unit does not turn on when external power is connected. Green
external power light does not illuminate.
1. Check power sources and connections.
2. Ensure wall outlet is live by plugging in a lamp.
Pump runs, but there is no suction.
1. Check that all tubing is connected properly.
2. Check tubing connections for breaks or leaks.
3. Ensure that float shut-off in collection container is not activated or filter cartridge occluded.
4. Check for leaks or cracks in collection container assembly.
Low suction.
1. Use vacuum regulator knob to increase suction level.
2. Check system for leaks.
Unit does not turn on (no external power is connected). 7314P
Series Only
1. Check that battery is fully charged and/or charge battery.
Battery will not charge (external power and charge indicator lights
should be illuminated during charge mode) 7314P Series Only
1. Verify that both external power and charge indicator lights illuminate.
2. Check power sources and connections.
3. Ensure wall outlet is live by plugging in a lamp.