16. CLEANING PROCEDURE
• Always disconnect the cot from the main power supply prior to cleaning.
To aid cleaning, the mattress platform panels can be removed. It is also advisable to remove
any accessories that are fastened to the cot.
These instructions apply to all accessories apart from soft products (e.g. mattresses).
• All surfaces to be wiped down with a disposable soft cloth moistened
with a mild detergent and diluted in warm water (40°C).
• The cot should be cleaned by starting with the cleanest parts of the cot and
systematically moving to the dirtiest parts. Extra care should be taken around
areas where excess dirt or dust may gather.
• The cleaning cloth should be changed during the cleaning process if it becomes soiled.
• Rinse down with clean water to remove detergent residue.
• Wipe surfaces down with 1,000 parts per million chlorine solution (0.1%).
• Dry off with a paper towel.
• Always ensure the cleaned parts are allowed to dry before putting the mattress
back in place.
Note: If any of the 3 stages stated above (detergent, rinse down & chlorine solution) are
omitted or combined it will reduce the effectiveness of the clean.
In cases of blood spills or other bodily fluids it is recommended that a chlorine solution of
10,000 parts per million (1%) is used instead.
Note: The use of neat bleach or similar surface cleaners is not recommended as damage
may be caused to the cot surfaces.
When removing the mattress panels for cleaning, take care when cleaning around fixing
points and over areas which could pose a sharp edge.
16.1 Cleaning & Disinfection Guidelines