ME3000 Analog Electronics        

                          

Lab 6 - 6/19  

 

                     

4. 

Voltage Gain Estimation 

21.  Unplug the jumpers from J28 and J29. 
22.  Set R

L

 = 470 

 by connecting jumpers to J30 and J31. 

23.  Set the function generator output to 40 mVpp and the output frequency to the resonance 

frequency value measured earlier. 

24.  Connect the function generator output to TP1 and reference to GND. 
25.  Connect the oscilloscope CH1 to TP1 and GND. 
26.  Connect the oscilloscope CH2 to TP6 and GND. 
27.  Set the trigger signal to CH1. 
28.  Record the peak-to-peak voltage at the input (CH 1) and output (CH 2), and determine the 

magnitude of the voltage gain, A

V

29.  Determine the amplifier voltage gain using 

)

||

(

L

ce

m

V

R

r

g

A

and the r

ce

 and g

m

 values 

calculated earlier. Compare the calculated value with the value measured in step 28. 

30.  Repeat steps 28 and 29 for R

L

 = 1470 

 by unplugging the jumpers from J30 and J31 and 

connecting the jumpers to J28 and J32. 

31.  Turn off the power supply and disconnect all the cables from the Class A Tuned Amplifier 

section. 

Содержание ME3000 Analog Electronics Lab 6

Страница 1: ...signing an RF tuned amplifier ii To demonstrate AC measurements on a Class A amplifier Equipment Required i ME3000 M2 Analog Electronics Training Kit ii Digital Multimeter recommendation Agilent 34405...

Страница 2: ...gion is almost linear Two types of amplifiers can be designed tuned and untuned amplifiers An untuned amplifier has a wider bandwidth than a tuned amplifier as shown in Figure 1 A tuned amplifier cons...

Страница 3: ...between 1 MHz to 300 MHz Great care must be taken in designing an RF amplifier as most of the components will not portray ideal characteristics at radio frequencies Most of the RF Class A amplifiers...

Страница 4: ...etails 5 Use the multimeter to measure VCC applied to the transistor between terminals TP4 and GND Calculate the voltage at the transistor base using the resistive voltage divider formula CC B V R R R...

Страница 5: ...to 8 MHz in 100 kHz increments 17 Record the peak to peak voltage at the input CH 1 and output CH 2 and determine the magnitude of the voltage gain AV as a function of frequency 18 Plot the AV versus...

Страница 6: ...25 Connect the oscilloscope CH1 to TP1 and GND 26 Connect the oscilloscope CH2 to TP6 and GND 27 Set the trigger signal to CH1 28 Record the peak to peak voltage at the input CH 1 and output CH 2 and...

Страница 7: ...urrent limit to 0 5 A Next set the supply to 25 V positive supply Set the voltage to 15 V and current limit to 0 5 A Repeat the same procedure for 25 V negative supply to set the voltage to 15 V and c...

Страница 8: ...e U2741A there are two LED indicators The power indicator lights up once the U2741A is powered on There is a system error if the indicator blinks after the U2741A is powered on The USB indicator will...

Страница 9: ...make a DC voltage measurement Measuring DC Current 1 The DC current measurement function of the U2741A has the following features three ranges to select 10 mA 100 mA 1 A and 2 A or auto range input i...

Страница 10: ...he output of the function generator after the desired parameters are set 14 The U2761A is able to output five standard waveforms that are Sine Square Ramp Triangle Pulse and DC 15 You can select one o...

Страница 11: ...ch panel of the interface Table A 2 Features of the AMM Function Generator Interface No Panel Features 1 Waveform Pattern Selection Select various types of output function by clicking the buttons 2 Wa...

Страница 12: ...for Sine wave in Vrms Duty Cycle Limitations For Square waveforms the U2761A may not be able to use the full range of duty cycle values at higher frequencies as shown below 20 to 80 frequency 10 MHz 4...

Страница 13: ...to half You cannot specify the output amplitude in dBm if the output termination is currently set to high impedance The units are automatically converted to Vpp You may configure the output terminati...

Страница 14: ...he oscilloscope to the PC using a USB cable 2 Power on the oscilloscope 3 Launch the Agilent IO Control and the Agilent Measurement Manager AMM 4 The Select USB Device dialog box will appear displayin...

Страница 15: ...quisition tab Displays the time domain waveform for the oscilloscope 3 FFT Analysis tab Displays the FFT spectrum of the signal 4 Configuration summary Displays the configured functions and settings 5...

Страница 16: ...rom channel 1 channel 2 click 1 2 or press the shortcut key F5 F6 2 To toggle the channel on or off click the channel buttons on the vertical control panel or click the toolbar to toggle the channel o...

Страница 17: ...orizontal scale causes the waveform to expand or contract in the center of the screen It provides functions of Time Base Delay and Mode for the horizontal scale adjustment This is shown in Figure B 87...

Страница 18: ...n is located on the toolbar of the soft front panel 2 Click to activate the automated measurement and cursor system This will display the window as shown below Figure B 11 Soft Front Panel of the Meas...

Страница 19: ...Scale and Run Stop 1 AutoScale automatically configures the oscilloscope to best display the input signal by analyzing any waveforms connected to the channel and external trigger inputs If AutoScale f...

Отзывы: