I. Descriptions   

Kjeldahl  method  is  a  classic  method  for  nitrogen  determination.  Currently,  the  Kjeldahl  method  is 

frequently  applied  for  determining  the  nitrogen  content  of  the  soil,  foods,  agricultural  products,  feeds  and 

other  substances.  When  this  method  is  applied  for  the  sample  determination,  three  processes  including 

digestion,  distillation,  and  titration  will  be  gone  through,  and  the  distillation  is  an  important  process  in 

Kjeldahl method.     

ATN-300  Automatic  Kjeldahl  Apparatus  is  a  kind  of  automatic  system  researched  and  development 

based on the  classic  Kjeldahl method,  which  is  featured by  user-friendly  operation and easy  learning  due to 

application of the Chinese menu and man-machine interface. 

With two parts of the manual distillation and automatic distillation, ATN-300 is available for manual and 

automatic  water  filling,  alkali  filling,  acid  filling,  and  distillation,  which  brings  large  convenience  to  the 

laboratory personnel. 

The  automatic  part  of  ATN-300  can  save  ten  kinds  of  different  combinations  for  water  filling,  alkali 

filling, acid filling, and distillation time for setting and application by the user.   

ATN-300 has applied the British self-sucking pump method for liquid feeding rather than the inflatable 

method applied by many domestic manufacturers, which can place the liquid feeding barrel under the control 

console and select the liquid feeding barrel with large capacity. 

ATN-300  has  been  furnished  with  numerous  protection  measures,  such  as  the  distillation  tube  absence 

protection (no heating or distillation without distillation tube); steam generator overpressure alarm protection 

(shutdown  and  alarm  in  case  of  abnormal  pressure);  steam  generator  water  shortage  protection,  and  earth 

leakage protection, so as to ensure safety of operators. 

 

II. Working Principles   

When all  samples  are  digested,  the  distillation  can  be  performed  after  having  been cooled  down  to  the 

room temperature. The principles of the distillation process are as follows:   

(NH

4

)

2

SO

4

+NaOH=Na

2

SO

4

+2H

2

O+2NH

3

↑ 

Under the alkalic conditions, separate the ammonia with the high temperature steam; after that, collect it 

with  the  conical  flask  containing  boric  acid  absorption  liquid  (containing  mixed  indicator);  lastly,  perform 

Содержание DW-ATN-300

Страница 1: ...N 300 Drawell International Technology Limited Shanghai Drawell Scientific Instrument Co Ltd Add Suite 1506 Lane581 XiuChuan Rd PuDong New Area Shanghai China Tel Fax 0086 21 50966080 Web www drawell com cn Email sales01 drawell com cn ...

Страница 2: ...s Structure 6 V Installation Method 7 5 1 Pre installation Check 7 5 2 Installation Conditions 7 5 3 Installation Steps 7 5 4 Debugging 7 VI Application Methods and Steps 8 6 3 Flow Calibration 9 6 4 Program Settings 9 6 5 Manual Distillation 10 6 6 Automatic Distillation 11 VII Reagent Preparation Method 12 VIII Precautions for Use of Apparatus 13 ...

Страница 3: ...ctions Danger Property or personal injuries may be caused please use the machine as required Special power supply Reliable grounding Flammable and explosive No transformation 1 Please use the power supply and voltage conforming to the product requirements 2 The applicable voltage for the equipment is AC220V 50HZ 3 The product shall be reliably grounded and kept away from electromagnetic interferen...

Страница 4: ...7 The working environment shall be free from dust and flammable and explosive gases 8 The working environment shall be of 10 45 and relative humidity less than or equal to 80 9 The apparatus shall be installed in the place without direct sunlight 10 As the reagent and other solution are corrosive safety appliances shall be equipped during operation for safety ...

Страница 5: ...y Card 1 Product Qualification Certificate 1 Accessories in the accessory box Distilled water inlet pipe 6mm 1 5m 1 Alkali liquid inlet pipe 6mm 1 5m 1 Boric acid solution inlet pipe 6mm 1 5m 1 Waste distilled water pipe 6mm 1 5m 1 Condensed water inlet and outlet pipes 8mm 2 5m 1 for each Digestion pipe 40 300mm 1 Gloves 1pair Power line 1 8m 1 ...

Страница 6: ... combinations for water filling alkali filling acid filling and distillation time for setting and application by the user ATN 300 has applied the British self sucking pump method for liquid feeding rather than the inflatable method applied by many domestic manufacturers which can place the liquid feeding barrel under the control console and select the liquid feeding barrel with large capacity ATN ...

Страница 7: ...water consumption 1 2L 2 4L min 3 5 Amounts of water filling alkali filling acid filling 0 99ml 3 6 Sample weight Solid 6g liquid 30mL 3 7 Operation mode Dual modes of manual distillation automatic distillation 3 8 Automatic distillation mode Ten kinds of distillation programs 3 9 Alarm and shutdown temperature 120 3 10 Power voltage AC220V 50HZ 3 11 Maximum power 1600w 3 12 Maximum current 7 5A 3...

Страница 8: ...wer switch 6 Distillate outlet pipe boric acid filling pipe 7 Conical flask 8 Digestion pipe tray 9 Acid liquid barrel 10 Alkali liquid barrel 11 Distilled water barrel Figure of the Back side Structure of the Apparatus see Figure 4 2 1 Distilled water inlet 2 Alkali liquid inlet 3 Boric acid solution inlet 4 Waste liquid outlet 5 Condensed water outlet 6 Condensed water inlet 7 Power socket The m...

Страница 9: ...8 C 6 The drain tank shall be 10cm lower than the apparatus outlet to ensure a smooth drainage 7 As the equipment is a kind of precise apparatus acute vibration or transverse and upside down placement shall be avoided 5 3 Installation Steps refer to Figure 4 1 and Figure 4 2 1 Place the apparatus on a stable experiment table 2 Connect one end of the condensed water inlet pipe with the condensed wa...

Страница 10: ...rtup is shown in the Figure 6 1 Function selection 6 2 Press key to enter the function selection page See Figure 6 2 Figure 6 2 Function selection F1 Correction Flow correction F2 Manual Manual distillation F3 Automatic Automatic distillation F4 Program Program setting 6 3 Flow Calibration Prior to flow calibration connect with the container after that press F6 F7 F8 keys for several times to fill...

Страница 11: ...the operation is ended record the distilled water volume in the graduated cylinder C Boric acid flow calibration Place the distillate outlet pipe into the graduated cylinder press F7 key to select the acid flow calibration After the operation is ended record the boric acid solution in the graduated cylinder D Press SET after above steps are completed E Input the recorded volume mL value into the c...

Страница 12: ...li 12ml Alkali filling amount 12 ml Acid 12ml Acid filling amount 12 ml Time 123s Distillation time 123 S A Press SET key to perform program parameter settings B After each setting press ENT once for confirmation using the following sequence C SET Program X selection ENT Water filling amount ml ENT Alkali filling amount ml ENT Acid filling amount ml ENT Distillation time S ENT ALM Remember After p...

Страница 13: ...rement press F3 again to stop the distillation D Press F4 to select the water filling when the solution in the digestion pipe reaches the requirement press F4 again to stop the water filling E Press F5 drainage means to drain out water in the steam generator When impurities occur in the steam generator or the apparatus is not used for a long time water in the steam generator shall be changed or dr...

Страница 14: ...olume such as 100ml may be different the users are recommended to perform trial distillation for several times according to the actual conditions to master set the distillation duration VII Reagent Preparation Method 7 1 Prepare 2L 30 40 sodium hydroxide solution and pour it into the alkali barrel suggestion The 35 sodium hydroxide solution is recommended for it will not easily crystallize and blo...

Страница 15: ...k the rubber plug 8 4 As the digestion pipe will be very hot after distillation gloves shall be worn upon taking it down to avoid scalding 8 5 The condensed water must be opened during distillation 8 6 The filled alkali must ensure the alkalinity of the digestion solution greater than or equaling to 11 8 7 If the apparatus is not used for a long time in winter water in the steam generator must be ...

Страница 16: ...rnational Technology Limited Shanghai Drawell Scientific Instrument Co Ltd Add Suite 1506 Lane581 XiuChuan Rd PuDong New Area Shanghai China Tel Fax 0086 21 50966080 Web www drawell com cn Email sales01 drawell com cn ...

Отзывы: