66
Maintenance
en
4.5.8
Performing the high pressure leak test (optional)
1. Set the selector switch on the test unit to
.
2. Open cylinder valve.
3. Note the cylinder pressure displayed on the Bodyguard.
If the cylinder pressure is less than 165 bar, the high pressure
leak test cannot be performed. In this case, the compressed
oxygen cylinder must be filled.
If the cylinder pressure is greater than 165 bar, the "Close
cylinder valve" symbol appears on the Bodyguard display.
The alarm tone sounds twice.
4. Close cylinder valve.
On the Bodyguard display, the bar chart runs from left
to right.
The "Remaining usage period" symbol appears on the
Bodyguard display. A single alarm tone sounds. The bar chart
runs once again from left to right.
The result of the test is output after approx. 15 seconds.
The following results may occur:
4.5.9
Checking the constant metering valve and bypass
valve
1. Set the selector switch on the test unit to
until the
protective cap fits over the pressure relief valve tappet.
2. Set the selector switch on the test unit to
.
3. Pull the open side of the protective cap over the pressure
relief valve tappet and hold. Pump until the protective cap is
secured by the filled breathing bag.
4. Open cylinder valve.
5. Briefly press the red button for the bypass valve.
Oxygen must be heard to flow into the circuit (flow noise).
6. Pump until +10 mbar is reached.
7. Set the selector switch on the test unit to
l/min
.
After a short settling time, the flow must be between 1.5 and
1.9 l/min
1)
.
4.5.10 Checking the minimum valve
1. Set the selector switch on the test unit to
and
remove the protective cap.
2. Wait until flow noises can be heard on the minimum valve.
The opening pressure must be b0.1 mbar and
+4.5 mbar.
4.5.11 Checking the residual pressure warning
1. Set the selector switch on the test unit to
.
2. Close cylinder valve.
3. Observe the Bodyguard.
The first residual pressure warning should be issued at
approx. 55 bar.
The alarm sounds intermittently. The red LEDs flash.
4. Remove the connecting piece from the test unit connector.
The device is vented automatically.
5. Fit the protective cap onto the connecting piece.
4.5.12 Switching off the Bodyguard and checking the
battery charge status
1. Simultaneously hold the right and left hand buttons on the
Bodyguard pressed down until a short beep sounds.
2. Release the buttons.
The charge status of the battery is displayed for 3 seconds.
The Bodyguard then switches off.
The charge status is displayed as follows:
Result
next steps
An error has occurred.
The "X" symbol appears on
the Bodyguard display.
The alarm sounds four times.
Do not use the closed-circuit
breathing apparatus.
Check once again that the
closed-circuit breathing
apparatus is functioning
correctly or have the de-
vice serviced by Dräger or
trained experts.
The closed-circuit breathing
apparatus is OK.
The "Open cylinder valve"
symbol appears on the Body-
guard display. The alarm
tone sounds twice.
Open cylinder valve.
1) Only applies at a cylinder pressure of 180 to 200 bar.
NOTICE
The second remaining pressure warning cannot be
tested with this test.
To determine the exact trigger values, a pressure
gauge comparative measurement must be carried out.
This is described in the apparatus maintenance
manual.
Display on the Bodyguard
Meaning
Battery OK:
The battery charge is ade-
quate, the battery need not
be replaced.
Battery warning 1:
A long warning tone sounds
at the same time.
With normal use of the Body-
guard, the closed-circuit
breathing apparatus remains
fully functional for up to
4 hours. To ensure safe us-
age, however, the battery
should be replaced when this
warning is issued during the
self-test before use.
Battery warning 2:
At the same time, 5 short
alarm tones sound and the
red LEDs flash.
The Bodyguard automatically
powers off and any further
operation is impossible.
The battery must be replaced
before before any new opera-
tion.
i
i
33621851.eps
bar
20221851.eps
bar
20321851.eps
bar
Содержание PSS BG 4 RP
Страница 1: ...Dr ger PSS BG 4 RP i ru 4 en Instructions for Use 40...
Страница 2: ......
Страница 3: ...9 2 1 3 4 5 6 7 8 10 11 12 14 16 17 18 19 20 21 13 15 1 2 3 4 5 6 7 1 2 3 4 5 5 7 8 33321851 eps B A 6...
Страница 9: ...9 ru 2 7 2 7 1 Dr ger Dr ger Dr ger Dr ger Dr ger Dr ger i i...
Страница 12: ...12 ru Dr ger FPS 7000 RP Panorama Nova RP 1 2 2 7 4 PCas APECas 1 2 i i...
Страница 20: ...20 ru 4 1 6 5 6 X X1 X2 X X 3 X CO2 X X4 CO2 X X4 X5 1 2 3 Dr ger 4 CO2 4 5 27 5...
Страница 28: ...28 ru 4 5 1 1 3 2 3 3 3 1 15 4 4 5 2 1 2 3 4 1 4 Bodyguard 1 2 3 4 5 6 7 OK 34921851 eps L min 6 5 3 4 7 1 2...
Страница 35: ...35 ru 5 6 CO2 CO2 7 Dr ger CO2 Dr ger CO2 1 CO2 2 CO2 Dr ger Dr ger CO2 CO2...
Страница 38: ...38 ru PC PCas TR...
Страница 74: ......
Страница 75: ......
Страница 76: ......
Страница 77: ......