Description

49

en

After use

Uncouple the closed-circuit breathing apparatus: Press 
button and at the same time pull out the plug-in coupling 
from the connection piece.

2.7.5

Cleaning and disinfecting

The mask must be cleaned and disinfected each time it is 
used.

Clean and disinfect the full face mask as follows:

Before cleaning, remove accessories and drinking valve 
mouthpiece from the mask and clean them separately.

If a Dräger MS-COM was fitted, attach the protective cap 
onto the Dräger MS-COM adapter to avoid injury.

If the mask has a drinking valve, the valve must be cleaned 
as described in "Clean and disinfect" drinking valve".

Position the wiper in the middle of the visor.

Clean all parts with lukewarm water and Sekusept

®

 cleaner 

using a soft cloth (temperature: max. 30 °C, concentration: 
0.5 - 1 %)

1)

.

Thoroughly rinse all parts under running water.

Prepare a disinfection bath consisting of water and Incidin

® 

Rapid (temperature: max. 30 °C, concentration: 1.5 %)

2)

.

Place all parts to be disinfected into the disinfectant bath 
(duration: 15 minutes).

Rinse all parts thoroughly under running water.

Allow all parts to dry in the open air or in a drying closet 
(temperature: max. 60 °C). Keep out of direct sunlight.

Fit the drinking valve mouthpiece and accessories.

Cleaning and Disinfecting the Drinking Valve

Prepare the cleaning solution (completely dissolve 1 
cleaning tablet in 1 litre of drinking water) and fill it into the 
drinking reservoir.

Replace the mouthpiece with the drain hose.

Connect the drinking reservoir to the outside of the drinking 
valve with a hose.

Squeeze the cleaning solution from the reservoir through 
the drinking valve until a small amount of the solution 
emerges from the drain hose.

Close off the drain hose with a hose clip.

Allow the cleaning solution to act for 3 minutes.

Open the hose clip and squeeze roughly half of the 
cleaning solution out of the reservoir and through the 
drinking valve.

Close the tube clip again and let the cleaning solution act 
for a further 3 minutes.

Open the hose clip and squeeze the remaining cleaning 
solution through the drinking valve.

Fill the drinking reservoir with clear drinking water.

Squeeze the water out of the reservoir and through the 
drinking valve.

Disconnect the drinking reservoir from the hose.

Using a compressed air gun, blow oil-free compressed air 
into the hose until the drinking valve is dry.

Dismantle the hose and the drain hose.

CAUTION

To prevent damage, avoid pulling on the equipment 

connector!

Full-

face 

mask

Remove the helmet if necessary.

Push the sliding buckles of the neck straps 
forward with your thumb and loosen the straps.

Pull the mask away from your face over your 
head.

Mask/

helmet

combi-

nation 

with S-

fix 

adapter

Open the chin strap on the helmet.

Pull both mask/helmet adapters evenly to the 
rear, then flip them away from the helmet. 

Remove mask.

Remove the helmet.

Mask/

helmet

combi-

nation 

with Q-

fix 

adapter 

Open the chin strap on the helmet.

Pull both mask/helmet adapters forwards 
slightly and press to release the snap fasteners.

Pull both mask/helmet adapters evenly to the 
rear, then flip them away from the helmet.

Remove mask.

Remove the helmet.

WARNING

The mask body must be cleaned sufficiently 

thoroughly that no residues from skin care products 

remain on it. Otherwise the elasticity in the sealing 

area can be lost and the mask will no longer sit 

correctly.

If this warning is not heeded, poisoning or death can 

result.

CAUTION

Do not use any solvents (e.g. acetone, alcohol) or 

cleaning agents containing abrasive particles for 

cleaning and disinfecting. Only use the prescribed 

process and use the cleaning and disinfection agents 

given. Other agents, dosages and contact times may 

damage the product.

The undiluted agents are damaging to health if they 

come into direct contact with the eyes or skin. For this 

reason, wear safety goggles and protective gloves 

when working with these agents.

!

!

!

1) Sekusept

®

 is a registered trademark of Ecolab Deutschland 

GmbH Note the separate information on the use of the product.

2) Incidin

®

 is a registered trademark of Ecolab USA Inc. Note the 

separate information on the use of the product.

Mask-

helmet 

combi-

nation

Relubricate the sliding surfaces (ball joint, guide 
plate) of the mask/helmet adapter with Voltalev

®

 

or Molykote

®

111.

Содержание PSS BG 4 RP

Страница 1: ...Dr ger PSS BG 4 RP i ru 4 en Instructions for Use 40...

Страница 2: ......

Страница 3: ...9 2 1 3 4 5 6 7 8 10 11 12 14 16 17 18 19 20 21 13 15 1 2 3 4 5 6 7 1 2 3 4 5 5 7 8 33321851 eps B A 6...

Страница 4: ...20 4 2 21 4 2 1 21 4 2 2 21 4 2 3 21 4 2 4 CO2 21 4 2 5 22 4 2 6 22 4 2 7 22 4 2 8 23 4 2 9 Switch Box 23 4 2 10 23 4 2 11 24 4 3 24 4 4 24 4 4 1 25 4 4 2 25 4 4 3 25 4 4 4 25 4 4 5 25 4 4 6 25 4 4 7...

Страница 5: ...5 ru 1 1 1 Dr ger Dr ger Dr ger Dr ger Dr ger Dr ger 1 2 2 2 1 2 1 1 1 2 3 4 Bodyguard 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 CO2 B 1 2 3 4 5 SwitchBox 6 7 i i...

Страница 6: ...1 2 3 4 5 6 7 8 2 2 2 2 1 PSS BG 4 RP CO2 T 15 C 40 C 10 C 2 2 2 SwitchBox Bodyguard SwitchBox Bodyguard 1 2 3 Bodyguard 50 60 75 Bodyguard 30 95 Bodyguard Bodyguard Bodyguard Bodyguard 2 6 8 25 10 3...

Страница 7: ...7 ru Bodyguard 5 Bodyguard Windows Bodyguard Bodyguard 2 3 4 2 4 4 180 0 C 15 C 40 C 72 0 C 10 C 25 C CO2 i i 0 C 6 C X 6 C 15 C X X 15 C 40 C X X X...

Страница 8: ...259 2009 EC 2016 425 TP TC 019 2011 www draeger com product certificates PSS BG 4 RP Bodyguard 2 6 Bodyguard Bodyguard Bodyguard PSS BG 4 RP TP TC 019 2011 Ta 40 C to 60 C PO ExiaI X O ExiaIICT4 X X D...

Страница 9: ...9 ru 2 7 2 7 1 Dr ger Dr ger Dr ger Dr ger Dr ger Dr ger i i...

Страница 10: ...000 RP A1 A2 A3 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 Panorama Nova RP A4 1 2 3 4 5 6 7 8 9 10 A1 12 11 10 14 13 1 6 8 9 5 4 3 2 7 19 20 21 A3 00121767 eps 1 2 3 4 5 7 8 10 6 9 A4 A2 1...

Страница 11: ...7000RP Dr ger FPS 7000 RP Dr ger Panorama Nova RP Dr ger MS COM 40 o C 60 o C Dr ger FPS 7000 RP Dr ger Panorama Nova RP Dr ger PSS BG 4 RP 2 7 3 89 656 EEC BGR 190 AS NZS 1715 2009 S M L 1 2 3 M L S...

Страница 12: ...12 ru Dr ger FPS 7000 RP Panorama Nova RP 1 2 2 7 4 PCas APECas 1 2 i i...

Страница 13: ...000 RP Panorama Nova RP Dr ger 2 7 5 Dr ger COM Dr ger COM Sekusept Cleaner 30 C 0 5 1 1 ncidin Rapid 30 C 1 5 2 15 60 C 1 1 S fix Q fix 1 Sekusept Ecolab Deutschland GmbH 2 Incidin Ecolab USA Inc Vol...

Страница 14: ...14 ru 3 3 2 7 6 BGR 190 2 7 7 15 o C 25 o C ISO 2230 3 3 1 CO2 6 CO2 6 CO2 500 4 5 27 24 CO2 0 C 15 C 3 2 1 2 4 5 27 3 3 1 3 3 1 15 2 CO2 4 4 8 26 6 2 6 X X X1 1 2 X X2 2 6 X X X CO2 4 5 27 i i...

Страница 15: ...15 ru 3 3 3 2 15 4 3 3 3 16 5 3 3 4 16 6 3 3 1 15 7 3 3 5 17 3 3 1 1 2 3 1 2 3 3 3 2 99 5 O2 0 05 N2 Ar 5000 ppm CO 5 ppm CO2 300 ppm 2 5 2 200 1 2 30321851 eps 30421851 eps 100mm 108mm 34621851 eps...

Страница 16: ...16 ru 3 3 4 2 5 1 6 3 3 3 1 40 2 15 C 16 3 8 4 1 3 4 2 3 4 5 1 2 6 7 3 3 4 1 10 Bodyguard 4 8 34 2 3 4 Bodyguard 0 5 34721851 eps 30821851 eps 2 1 3 30521851 eps 30621851 eps 1 2 4 3...

Страница 17: ...7 ru Bodyguard 3 3 5 1 2 3 4 5 A 6 B 7 8 9 10 11 12 13 Bodyguard Dr ger 1 Bodyguard 4 2 5 Bodyguard 20021851 eps psi bar 20121851 eps bar 20221851 eps bar 20321851 eps bar 30021851 eps 30121851 eps B...

Страница 18: ...18 ru 14 15 16 Bodyguard 4 8 34 17 Bodyguard 18 10 19 3 4 3 4 1 CO2 10 15 55 Bodyguard 5 Bodyguard C 3 4 2 2 6 8 180 i i 30721851 eps Dr ger...

Страница 19: ...19 ru 3 4 3 Bodyguard Bodyguard Bodyguard Bodyguard Bodyguard 3 5 1 2 3 4 5 Bodyguard Bodyguard 3 Bodyguard Bodyguard 6 a b c d 7 4 19 4 1 4 2 21 2 4 3 24 3 4 4 24 4 4 5 27...

Страница 20: ...20 ru 4 1 6 5 6 X X1 X2 X X 3 X CO2 X X4 CO2 X X4 X5 1 2 3 Dr ger 4 CO2 4 5 27 5...

Страница 21: ...2 4 2 4 21 6 4 2 5 22 7 4 2 6 22 8 4 2 7 22 9 4 2 8 23 10 Switch Box 4 2 8 23 11 4 2 10 23 4 2 1 1 2 2 3 3 1 4 5 1 4 2 2 1 2 3 4 4 2 3 1 2 3 4 5 4 2 4 CO2 1 CO2 1 2 2 Switch Box 4 2 11 24 3 DIN EN 12...

Страница 22: ...O2 7 CO2 8 4 2 5 1 1 2 2 3 4 4 5 3 6 5 6 Molykote 111 7 4 2 6 1 A 1 2 B 3 4 2 11 24 3 B 2 4 5 B 5 B 4 6 4 2 7 1 1 2 2 31021851 eps 3 4 i i 31121851 eps 1 2 31221851 eps 3 4 5 6 31521851 eps 3 4 2 5 B...

Страница 23: ...A 4 A 5 4 2 1 5 3 6 3 7 8 9 B 6 10 a 9 b 8 c 7 d e 10 f 4 2 9 Switch Box 1 Switch Box 2 Switch Box 4 2 10 1 1 2 3 2 90 4 Bodyguard 32321851 eps 1 2 3 6 31721851 eps A B 4 5 31921851 eps 7 8 9 10 31621...

Страница 24: ...5 15 6 7 60 C Dr ger R 34 098 4 4 1 4 4 1 25 2 4 4 2 25 3 4 4 3 25 4 4 4 4 25 5 4 4 5 25 6 4 4 6 25 7 Switch Box 4 4 7 25 8 CO2 4 4 8 26 9 4 4 9 26 10 4 4 10 27 11 3 3 2 15 12 3 3 1 15 1 R 33 777 320...

Страница 25: ...guard 3 90 4 4 4 2 1 2 4 4 3 1 DIN EN 12 021 2 3 1 4 4 4 1 A 1 2 3 4 5 B 4 6 A 2 A 3 4 4 5 1 2 4 4 6 1 2 A 1 3 A 2 A 3 4 B 4 4 4 7 Switch Box Switch Box 32421851 eps 1 i i 32521851 eps 4 B A 2 1 3 326...

Страница 26: ...8 CO2 1 CO2 2 CO2 5 3 CO2 6 4 a b 5 7 6 4 Molykote 111 7 CO2 8 2 1 Switch Box 3 9 CO2 10 4 4 9 1 2 1 3 3 Molykote 111 4 2 5 Molykote 111 CO2 CO2 32721851 eps 7 6 5 4 30221851 eps 1 2 3 i i 1 1 2 3 329...

Страница 27: ...27 ru 4 4 10 A 90 1 2 3 A A 1 4 A 2 A 3 5 A 5 CO2 6 A A 4 A 7 8 B 4 5 CO2 15 6 CO2 CO2 6 R 33 588 33021851 eps 33121851 eps B A 1 5 4 2 3 i i Dr ger Dr ger RZ 7000 39 Dr ger i i...

Страница 28: ...28 ru 4 5 1 1 3 2 3 3 3 1 15 4 4 5 2 1 2 3 4 1 4 Bodyguard 1 2 3 4 5 6 7 OK 34921851 eps L min 6 5 3 4 7 1 2...

Страница 29: ...3 1 2 10 3 10 4 2 3 21 4 5 4 1 2 3 10 4 10 4 2 3 21 4 5 5 1 2 3 10 25 4 5 6 1 2 7 0 5 3 1 1 4 5 7 1 2 3 7 9 Bodyguard Bodyguard 35021851 eps i i 35221851 eps L min i i Bodyguard Bodyguard 35121851 eps...

Страница 30: ...ru 4 5 8 1 2 3 Bodyguard 165 165 Bodyguard 2 4 Bodyguard 15 4 5 9 1 2 3 4 5 6 10 7 1 5 1 9 1 4 5 10 1 2 0 1 4 5 4 5 11 1 2 3 Bodyguard 55 4 5 Bodyguard Bodyguard X 4 Dr ger Bodyguard 2 i i 1 180 200 i...

Страница 31: ...31 ru 4 5 12 Bodyguard 1 Bodyguard 2 3 Bodyguard Bodyguard 1 Bodyguard 4 2 5 Bodyguard 33621851 eps bar 20221851 eps bar 20321851 eps bar...

Страница 32: ...ger Test It 6100 R50028 4 5 6 7 8 9 10 11 9 0 2 12 2 3 13 8 0 2 2 14 0 5 15 16 Dr ger RZ 7000 Dr ger Test It 6100 R50028 17 18 19 20 Bodyguard 4 5 12 31 21 3 3 1 15 1 2 3 4 5 6 7 8 Bodyguard OK i i 3...

Страница 33: ...Bodyguard 5 6 3 2 0 4 0 1 Bodyguard Bodyguard 7 Bodyguard 8 Bodyguard 4 7 Bodyguard Bodyguard 1 Bodyguard Bodyguard Bodyguard 0 0 2 Bodyguard Bodyguard 3 Bodyguard Bodyguard Bodyguard 90 21 716 33221...

Страница 34: ...a b 4 9 4 9 1 1 4 2 1 21 2 Switch Box 3 Switch Box 4 4 2 11 24 5 6 7 8 SW 17 9 4 9 2 1 2 3 SW 17 15 5 4 5 6 7 8 Switch Box 9 10 Switch Box 11 3 3 2 15 2 5 Bodyguard 165 Bodyguard Bodyguard 20321851 e...

Страница 35: ...35 ru 5 6 CO2 CO2 7 Dr ger CO2 Dr ger CO2 1 CO2 2 CO2 Dr ger Dr ger CO2 CO2...

Страница 36: ...Nova RP 13 8 x x 595 x 450 x 185 1 240 30 0 8 1 6 50 80 2 200 3 3 2 15 Bodyguard 200 10 40 0 5 6 C 15 C 20 5 C 4 C 20 5 C 40 C 72 Dr ger 40 C 240 1 60 C 120 1 90 C 86 1 40 C 240 1 60 C 40 1 90 C 25 1...

Страница 37: ...orama Nova EPDM PC RP R53306 Panorama Nova RP PC R53306 Panorama Nova Silicone RP PC R53316 1 R34800 CO2 R34360 CO2 R34277 CO2 R34367 R34633 Bodyguard 3355751 R33999 R34745 R34646 R34498 R34746 200 R3...

Страница 38: ...38 ru PC PCas TR...

Страница 39: ...39 ru 10 Dr ger RZ 7000 55 PSS BG 4 RP RZ 7000 Bodyguard 165 1 5 1 9 0 1 4 5 10 25 7 0 5 OK 1 7 9 1 4 10 10 53259 2009...

Страница 40: ...e and where applicable the water absorber 59 4 2 9 Removing the switch box 59 4 2 10 Disassembling waist belt and shoulder straps 59 4 2 11 Unfastening the plug in connectors on the medium pressure ho...

Страница 41: ...text that require greater awareness by the user A definition of the meaning of each icon is as follows 2 Description 2 1 Product overview 2 1 1 Closed circuit breathing apparatus Illustration on the f...

Страница 42: ...control line The sensor unit measures the available oxygen pressure in the compressed oxygen cylinder and the pressure in the breathing circuit The values are compared in the switch box Alarms are tr...

Страница 43: ...erefore always be maintained according to the respective instructions for use The cylinder pressure before use If the cylinder is not filled to capacity usage time is reduced proportionately The cylin...

Страница 44: ...d marking PSS BG 4 RP TR CU 019 2011 Ta 40 C to 60 C PO ExiaI X O ExiaIICT4 X Symbol Meaning Test completed successfully An error has occurred The error is indicated by the letter in front of the X Fu...

Страница 45: ...not modify the product Notify Dr ger in the event of any product or product component fault or failure Use in areas subject to explosion hazards Devices or components for use in explosion hazard areas...

Страница 46: ...alve 10 Drinking valve with mouthpiece 11 Cap 12 Mask body 13 Neck strap 14 Sliding buckle 15 Helmet 16 Mask helmet adapter 17 Chin strap 18 Mask fastening plate 19 Speech diaphragm 20 O ring 21 Clip...

Страница 47: ...otective equipment must be suitable for the respective workplace conditions Selecting the Correct Mask Size The mask bodies are available in sizes S M L inner masks in sizes 1 2 3 and head harnesses i...

Страница 48: ...e substances may be absorbed by the material of the Dr ger FPS 7000 RP and Panorama Nova RP and may possibly diffuse into them Further information can be obtained from Dr ger Potentially Explosive Are...

Страница 49: ...the remaining cleaning solution through the drinking valve Fill the drinking reservoir with clear drinking water Squeeze the water out of the reservoir and through the drinking valve Disconnect the d...

Страница 50: ...her the remaining oxygen supply is adequate for use The cylinder valve is fully closed after every use This means that the breathing circuit is vented automatically The connecting piece is sealed with...

Страница 51: ...only be filled with oxygen of the following quality characteristics Purity 99 5 vol O2 Water content 0 05 mg L oxygen Tasteless and odourless National regulations apply for the maximum permissible con...

Страница 52: ...over Ensure that the cross bar is fully open to avoid damage to the positioning aids on the breathing air cooler when the breathing air cooler is sealed with the cover 5 Align the positioning aids on...

Страница 53: ...the hose support centrally on the shoulder The other reinforcing rings should also rest on the shoulder 12 If the closed circuit breathing apparatus has a connecting piece with condensate trap ensure...

Страница 54: ...Bodyguard briefly Instead of the remaining period of use the temperature is displayed in C The display changes back automatically to the remaining period of use 3 4 2 In an emergency Press the yellow...

Страница 55: ...yguard switches off 6 Take off the closed circuit breathing apparatus a Open the waist belt Pinch the latch together Open the hook and eye b Draw the breathing hoses back over your head and lay them o...

Страница 56: ...ection by approved technical authority X5 1 Applies to closed circuit breathing apparatus that is constantly in use 2 Applies to reserve stocks 3 Only performed by expert personnel Send the pressure r...

Страница 57: ...he closed circuit breathing apparatus anti clockwise until they disengage 3 Take the breathing hoses off the closed circuit breathing apparatus 4 Drain off the condensate that has accumulated in the b...

Страница 58: ...te 111 thinly to the O ring 7 Re fit the pressure relief valve 4 2 6 Removing the breathing air cooler 1 Unhook the breathing bag from the breathing air cooler To do this stretch the tab on the sleeve...

Страница 59: ...cover 9 b Remove the valve disc 8 check and replace with a new valve disc if necessary c Ensure that the sinter filter 7 is fitted d Insert the shaft of the valve disc into the hole e Position the sp...

Страница 60: ...he constant metering valve see chapter 4 4 6 on page 61 7 Fit the switch box see chapter 4 4 7 on page 61 8 Fit the CO2 absorber see chapter 4 4 8 on page 61 9 Fit the connecting piece see chapter 4 4...

Страница 61: ...hose in to the pressure reducer B 4 It must engage into place Pull gently on the plugs to check that the medium pressure hose is sitting securely 4 4 7 Fitting the switch box Ensure that the cables ar...

Страница 62: ...g on the valve disk into the central hole and pull it until under cut becomes visible 2 Insert both valves 1 into the connecting piece They have different diameters and therefore cannot be confused Th...

Страница 63: ...tightness and functional test before initial commissioning and after every maintenance operation If the CO2 absorber is installed in the closed circuit breathing apparatus the functional test can be...

Страница 64: ...iece and insert the connecting piece into the connection for the closed circuit breathing apparatus 2 Set the selector switch on the test unit to 3 Pump slowly 4 Watch the test unit display The low pr...

Страница 65: ...4 5 6 Performing the leak test with positive pressure 1 Set the selector switch on the test unit to 2 Reduce the pressure on the test unit using the venting switch to approx 7 0 5 mbar and wait until...

Страница 66: ...e first residual pressure warning should be issued at approx 55 bar The alarm sounds intermittently The red LEDs flash 4 Remove the connecting piece from the test unit connector The device is vented a...

Страница 67: ...ctor switch on the test unit to and wait 2 to 3 minutes until the pressure in the system has stabilised 13 Then reduce the pressure in the test unit with the venting switch to approx 8 0 2 mbar and wa...

Страница 68: ...e battery If the Bodyguard is OK a pressure of 0 bar and a remaining period of use of 0 minutes is displayed The red LEDs flash 2 Switch off Bodyguard Simultaneously press the right and left hand butt...

Страница 69: ...sensor using an open end wrench SW 17 and remove from the pressure reducer 9 Insert the bypass valve switch and spring into the pressure reducer and screw the crown nut onto the pressure reduc er 4 9...

Страница 70: ...when the product is returned For information please contact the national sales organisations and Dr ger Alternatively the CO2 absorber can be disposed of as follows 1 Pack the CO2 absorber in a bag a...

Страница 71: ...ar 0 bar 5 bar Ambient conditions During operation Minimum temperature 6 C 15 C is permitted if the closed circuit breathing apparatus has been stored at a temperature range of 20 5 C Minimum temperat...

Страница 72: ...e use cartridge R34360 CO2 absorber refillable cartridge grey emp ty R34277 CO2 absorber refillable cartridge transpar ent empty R34367 Water absorber for drain valve R34633 Approved battery for Bodyg...

Страница 73: ...flow noise between 0 1 and 4 5 mbar Drain Valve Insert protective cap on relief valve Pump to fill system opens between 10 and 25 mbar Leak Test Reduce pressure to 7 0 5 mbar Press OK button pressure...

Страница 74: ......

Страница 75: ......

Страница 76: ......

Страница 77: ......

Страница 78: ...A Revalstra e 1 23560 L beck Germany Tel 49 451 882 0 Fax 49 451 882 20 80 www draeger com 9021791 GA 1114 032 Dr ger Safety AG Co KGaA Edition 03 October 2017 Edition 01 May 2008 Subject to alteratio...

Отзывы: