Dräger Polytron 7000
Technical data
NOTICE
Cross-sensitivities
The sensor cannot be used for the measurement of oxygen if helium is present.
►
Do not use the sensor for oxygen measurements if helium is present.
CAUTION
Sensor fault
If the sensor is exposed to high concentrations of unsaturated hydrocarbons,
alcohol or hydrogen, this can lead to failure of the device.
►
Make sure that the sensor is not exposed to such concentrations over a long
period.
Stabilisation time:
5 x t
0...90
Humidity
5 to 95 % rh, non-condensing
Temperature
-40 to +60 °C
b65 °C
Storage
Pressure
No effect
Humidity
30 to 70 % rh, non-condensing
(only relevant if sensor packaging is open)
Temperature
0 to +40 °C
Duration
Storage of sensors is not anticipated. Sensors
should be used from the point of arrival. Remain-
ing life span = expected life span - storage period
Effect of ambient parameters
Zero-point
Sensitivity
Temperature
-40 to 65 °C
< ± 0.3 vol% O
2
< ± 0.3 vol% O
2
Pressure
< ± 0.1 vol% O
2
Relative deviation from dis-
play at 1013 hPa:
< 2 % of the measured value
/ 100 hPa
Humidity
No effect
Relative deviation from dis-
play at 50 % rh:
< 2.5 % of the measured
value
Reaction time
t
0...20
t
0...90
0 to 25 vol% O
2
≤
10 seconds
≤
30 seconds
Ambient parameters
Содержание Polytron 7000
Страница 1: ...Dr ger Polytron 7000 Approved as type P3U and type P3FB Instructions for use Dr ger Polytron 7000...
Страница 53: ...QVWUXFWLRQV IRU XVH _ Dr ger Polytron 7000 Annex 29968...
Страница 57: ...QVWUXFWLRQV IRU XVH _ Dr ger Polytron 7000 Annex 15 3 Declaration of conformity 29964...
Страница 59: ...QVWUXFWLRQV IRU XVH _ Dr ger Polytron 7000 Annex...