Dräger Polytron 7000
Device settings
9 Device
settings
9.1
Pump (only with pump module)
9.1.1
Adjusting the pump output
Equipment:
– Pump adapter (order no. 8320900)
– flowmeter
1. Select
Settings
>
Instrument
>
Pump
>
Pump flow
.
The 4-20 mA interface issues the maintenance signal.
"
Flowalarm woud
be disabled
.
Please usea Flowmeter
." is displayed.
If a relay module is installed, the fault relay is energised.
2. Connect pump adapter and flowmeter.
3. Select
Next
.
4. Adjust the pump output setting and check at the flowmeter. For a short response
time select a high pump throughput rate.
5. Select
Next
.
The flow thresholds are displayed.
Flow threshold for error: 0.3 L/min
Flow threshold for notice: 0.4 L/min
9.1.2
Displaying the pump runtime
– Select
Settings
>
Instrument
>
Pump
>
Operating time
.
9.2
Alarms (only with relay module)
9.2.1
Switching the alarm on or off
1. Select
Settings
>
Instrument
>
Alarm
>
Alarm on/off
.
29808
OK
+
L/min
enable
Alarm relay is switched on.