Dräger Polytron 7000
Technical data
14.2 Pump
module
14.3
Operating conditions with DrägerSensor O
2
and O
2
LS
14.3.1 DrägerSensor
O
2
(6809720)
The DrägerSensor O
2
(6809720) is an electrochemical two-electrode sensor for the
measurement of oxygen (O
2
) in the ambient air. The sensor measures the
O
2
-partial pressure. Changes in pressure have an effect on the measured value in
this process. For 1013 hPA without oxygen depletion the sensor measures
20.9 vol% O
2
.
Supply voltage (DC)
for throughput rate 0.5 L/min
12 V to 30 V at the gas detector
for throughput rate 1.0 L/min
16 V to 30 V at the gas detector
for throughput rate 1.5 L/min
20 V to 30 V at the gas detector
Power consumption
for throughput rate 0.5 L/min
<2 W
for throughput rate 1.0 L/min
<4 W
for throughput rate 1.5 L/min
<6 W
Settings
Throughput rate
Settings range
approx. 0.5 L/min to 1.5 L/min (approx. 30 % to
100 %)
Factory setting
0.5 L/min
Warning and alarm
Flow warning
0.4 L/min
Flow alarm
0.3 L/min
Interior diameter at the tube
connection
5 mm
Ambient conditions
Temperature (during opera-
tion)
0 ... 55 °C (32 ... 130 °F)
Temperature (during storage) -40 ... 70 °C (-40 ... 150 °F)
Pressure
700 ... 1300 hPa
Humidity
0 ... 100 % rh, non-condensing
Ambient parameters
Pressure
20.7 to 38.4 in. Hg (700 to 1300 hPa)
Humidity
10 to 95 % rh, non-condensing
Temperature
-5 to +40 °C
briefly -20 to +55 °C
Storage
Pressure
No effect
Содержание Polytron 7000
Страница 1: ...Dr ger Polytron 7000 Approved as type P3U and type P3FB Instructions for use Dr ger Polytron 7000...
Страница 53: ...QVWUXFWLRQV IRU XVH _ Dr ger Polytron 7000 Annex 29968...
Страница 57: ...QVWUXFWLRQV IRU XVH _ Dr ger Polytron 7000 Annex 15 3 Declaration of conformity 29964...
Страница 59: ...QVWUXFWLRQV IRU XVH _ Dr ger Polytron 7000 Annex...