Instructions for use
Polaris 600
83
Technical data
Recommended separation distances
from wireless communication devices
To ensure that the full functional integrity of this
device is not compromised, there must be a
separation distance of at least 1.0 m (3.3 ft)
between this device and wireless high-frequency
communication equipment.
Immunity against
Test level and required electromagnetic environ-
ment
Electrostatic discharge (ESD) (IEC 61000-4-2)
Contact discharge: ±8 kV
Air discharge: ±15 kV
Electrical fast transient disturbances (bursts)
(IEC 61000-4-4)
Power cable: ±2 kV
Longer signal input lines/signal output lines: ±1 kV
Impulse voltages (surges) (IEC 61000-4-5)
Voltage phase conductor – phase conductor: ±1 kV
Voltage phase conductor – protective ground conduc-
tor: ±2 kV
Magnetic fields at mains frequency
(IEC 61000-4-8)
50 Hz: 30 A/m
Voltage dips and short interruptions of the sup-
ply voltage (IEC 61000-4-11)
Voltage dips of 30 % to 100 %, 8.3 ms to 5 s, different
phase angles
Radiated high-frequency disturbances
(IEC 61000-4-3)
80 MHz to 2.7 GHz: 3 V/m
Conducted high-frequency disturbances
(IEC 61000-4-6)
150 kHz to 80 MHz: 3 V, ISM bands: 6 V
Electromagnetic fields in the vicinity of wireless
communication devices
Various frequencies from 385 MHz to 5785 MHz:
9 V/m to 28 V/m