Instructions for use
Polaris 600
13
Safety-related information
Direct view of the operating field
Cautious positioning of the arm system
Risk of the light head overheating
Compatibility of displays
Observe the maximum load of the
display mount
Read and observe the instructions for
use of the displays
OR server
WARNING
Risk of personal injury
A camera is not intended to be used as the
sole visualization medium during an
operation.
Make sure that the OR personnel have a
direct view of the operating field at all times.
CAUTION
Risk of personal injury and property damage
The stops on the arm system may become
damaged.
When positioning the swivel arms and the
attached devices, take care not to use force to
pull the arm system beyond the stops.
WARNING
Risk of property damage
The light head may overheat during
operation.
Do not cover the light head completely or
partially.
WARNING
Risk due to incompatible displays
The displays used in combination with a
Polaris light system must be approved for
use in the patient environment according to
IEC 60601-1.
WARNING
Risk of personal injury and property damage
The specification for the maximum load on
the display mount must not be exceeded.
Comply with the maximum load of the
display mount.
WARNING
Risk of device malfunction
To properly use the light system, read and
comply with the instructions for use of the
respective display.
WARNING
Risk of patient injury
The OR servers used must conform to Safety
Class B of the IEC 62304 standard.
WARNING
Risk of patient injury
The OR severs used must conform to the
IEC 60601-1:1988 (or later) or IEC 60950-1
standards.