Calibrating the gas detector

9

en

5.3

Carrying out a manual span calibration 

Prepare the test gas cylinder, connect the test gas cylinder 
to the calibration adapter and connect the calibration 
adapter to the gas detector.

1. Open the maintenance menu, see Section 5.1 on Page 8.
2. Press the [OK] button while the span calibration symbol is 

flashing. The configured test gas concentration is 
displayed. The test gas concentration can be used or 
adjusted according to the concentration in the test gas 
cylinder.

3. Press the [

] button to change the configured test gas 

concentration. The first position flashed. Press the [

button to change the value of the flashing position. Press 
the [OK] button to confirm the value. Repeat this procedure 
to define the next three values. After the last confirmation 
with the [OK] button, the test gas concentration is 
complete.

4. Open the test gas cylinder valve to let test gas flow over the 

sensor (flow: 0.5 L/min).

5. Wait until the displayed measured value is stable (after at 

least 120 seconds).

6. Press the [OK] button to start the span calibration. The 

concentration display flashes. 

7. As soon as the measured value displays a stable 

concentration, press the [OK] button to confirm the span 
calibration or the [

] button to cancel the span calibration.

Span calibration passed: A short double tone sounds, and 

OK / gas name

 is shown in an alternating display. Press 

the [OK] button to return to measuring mode.

Span calibration failed: A tone sounds.

– – –

 is displayed instead of the measured value. The error 

symbol and the span calibration symbol are displayed. In 
this case, repeat the span calibration.

To test the measured value configuration times, allow t90 
test gas to flow over the calibration adapter on the gas 
detector. Check the results in accordance with the 
information in the table (see the supplementary 
documentation, order no. 90 33 741) up to a reading of 
90 % of the final display.

For Pac 8500, the span calibration is carried out 
consecutively for the various gasses.

5.4

Carrying out manual calibration with the 
Dräger CC-Vision Basic

To calibrate the gas detector using the Dräger CC-Vision Basic 
PC software:
1. Connect the gas detector to a PC using the communication 

module.

2. The calibration can be carried out using the Dräger 

CC-Vision Basic PC software. For further information, see 
the Dräger CC-Vision Basic online help.

A calibration interval can be set using the “Configurable 
operating time” (in days) feature, see Section 7.1 on 
Page 13.

5.5

Carrying out an automatic calibration 
with the Dräger X-dock

The gas detector can be automatically calibrated using the 
Dräger X-dock, see the Dräger X-dock instructions for use.

Содержание Pac 6000

Страница 1: ...Dr ger Pac 6x00 8x00 Technical manual i...

Страница 2: ......

Страница 3: ...ration 7 5 Calibrating the gas detector 8 5 1 Opening the maintenance menu 8 5 2 Carrying out a manual fresh air calibration 8 5 3 Carrying out a manual span calibration 9 5 4 Carrying out manual cali...

Страница 4: ...and non earthed metal parts of the housing may store a combustible level of electrostatic charge Activities such as carrying the device in a bag or attached to a belt operating the button field or cle...

Страница 5: ...ormity are provided in the enclosed supplementary documentation order no 90 33 741 The name plate on the gas detector must not be concealed 1 Alarm LEDs 6 OK button 2 Operation signal D Light 7 button...

Страница 6: ...curacy After the gas detector has been switched on the current measured value is shown in the display Check whether the warning appears If it is displayed we recommend performing a bump test see Secti...

Страница 7: ...vely The information display closes after the device ID When the gas detector is on The following information can be displayed depending on the configuration Error codes Peak concentration Time weight...

Страница 8: ...the last confirmation with the OK button the entire password flashes 3 Press the OK button to confirm the password or the button to cancel the password 4 If the correct password is entered a signal to...

Страница 9: ...r the button to cancel the span calibration Span calibration passed A short double tone sounds and OK gas name is shown in an alternating display Press the OK button to return to measuring mode Span c...

Страница 10: ...Wrong software version Update the software 117 User parameters implausible Check and adjust the configuration of the user parameters 131 Gas detector faulty Contact the service of Dr ger 132 Paramete...

Страница 11: ...rror on Dr gerSensor XXS measuring channel 2 Carry out a calibration see Section 5 on Page 8 327 Gas detector faulty Contact the service of Dr ger 328 Compensation channel calibration error on Dr gerS...

Страница 12: ...n Page 8 351 Dr gerSensor XXS measuring channel 2 warming up warm up phase 1 Wait until warm up time is complete 352 Dr gerSensor XXS measuring channel 2 warming up warm up phase 2 Wait until warm up...

Страница 13: ...ing the Dr ger CC Vision Basic PC software 1 Use the communication module order no 83 18 587 to connect the gas detector to a PC or the Dr ger X dock maintenance station 2 The configuration can be car...

Страница 14: ...on Basic PC software as described Dr ger recommends signing on the sensor with the Dr ger CC Vision Basic PC software even if the sensor code is identical to the previous one 7 5 Replacing the battery...

Страница 15: ...iaphragm must be replaced due to the special diaphragm 8 Device settings Only trained and qualified personnel may change the device settings 8 1 Factory setting The factory settings may differ for cus...

Страница 16: ...Main alarm3 Instrument alarm 1 The intensity of the vibration depends on the temperature 2 After the initial battery pre alarm the battery s life span amounts to between 1 day and 2 weeks under normal...

Страница 17: ...centers as specified by the applicable regulations Ambient conditions during operation Temperature temperature depending on the sensor Up to 30 C 55 C briefly up to 1 h 40 C 55 C Humidity 10 90 r h n...

Страница 18: ...m Acknowledgeable Latching Alarm threshold A2 1 60 ppm 10 ppm 23 Vol 2 ppm Acknowledgeable Latching TWA threshold value 1 3 3 Not applicable for Pac 6000 30 ppm 10 ppm 1 ppm STEL threshold value 1 3 6...

Страница 19: ...1 100 ppm 0 2 ppm 20 ppm 50 ppm 1 ppm 1 Vol Acknowledgeable Latching TWA threshold value 1 20 ppm 0 1 ppm 1 9 ppm 25 ppm 0 5 ppm 0 5 Vol STEL threshold value 1 40 ppm 0 1 ppm 3 8 ppm 50 ppm 1 ppm 2 V...

Страница 20: ...A2 1 1 ppm 20 ppm 20 ppm 0 2 ppm 0 2 ppm Acknowledgeable Latching TWA threshold value 1 0 5 ppm 0 1 ppm 0 1 ppm STEL threshold value 1 0 5 ppm 0 1 ppm 0 1 ppm Mean value period 15 min 15 min 15 min Me...

Страница 21: ...60 ppm CO 23 O 2 60 CO 60 ppm Acknowledgeable O 2 CO Latching O 2 CO TWA threshold value 1 5 ppm H 2 S 30 ppm CO 30 ppm CO 30 ppm CO STEL threshold value 1 10 ppm H 2 S 60 ppm CO 60 ppm CO 60 ppm CO M...

Страница 22: ...2 No effect Ethane No effect No effect 0 2 Ethanol No effect No effect No effect Ethylene No effect No effect 1 Hydrogen 0 35 No effect 1 5 Hydrogen chloride No effect No effect No effect Hydrogen cy...

Страница 23: ......

Страница 24: ...ger Safety AG Co KGaA Revalstra e 1 23560 L beck Germany Tel 49 451 882 0 Fax 49 451 882 20 80 www draeger com 90 33 742 TH 4623 700 Dr ger Safety AG Co KGaA Edition 01 July 2017 Subject to alteratio...

Отзывы: