Troubleshooting

en

6

Troubleshooting

An error message is displayed in the event of instrument faults. The number that appears below the message is used for service 
functions. If the fault remains after turning the device on and off several times, contact the service of Dräger.

Error code

Cause

Remedy

101

Life span of the gas detector expired.

Use a new Pac gas detector.

102

The user's service life counter has elapsed.

Reset the service life counter with the Dräger 
CC-Vision Basic PC software.

103

Gas detector faulty.

Contact the service of Dräger.

104

Program code checksum error.

Contact the service of Dräger.

105

Bump test interval expired.

Carry out bump test, see see Section 4.2.3 on 
Page 6.

106

Calibration interval expired.

Carry out a span calibration.

107

Software error

Contact the service of Dräger.

109

General error. For example, menu functions cannot 
be carried out.

Check the configuration, view the error code (e.g. on 
the information display or with the Dräger CC-Vision 
Basic PC software)

111

Alarm element test failed: Alarm lamp.

Repeat the alarm element test with X-dock.

112

Alarm element test failed: Alarm horn.

Repeat the alarm element test with X-dock.

113

Alarm element test failed: Vibration motor.

Repeat the alarm element test with X-dock.

114

Parameter check failed.

Correct the parameters and repeat the test with 
X-dock.

115

Gas detector deactivated by X-dock.

Activate the gas detector with X-dock.

116

Wrong software version.

Update the software.

117

User parameters implausible.

Check and adjust the configuration of the user 
parameters.

131

Gas detector faulty.

Contact the service of Dräger.

132

Parameter checksum error.

Carry out a sensor inauguration using the Dräger 
CC-Vision Basic PC software.

133

Wrong software version for the options board

Update the software.

134

Hardware error.

Contact the service of Dräger.

201

No valid zero adjustment of DrägerSensor XXS 
measuring channel 1.

Carry out a fresh air calibration, see Section 5.2 on 
Page 8.

202

No valid span calibration of DrägerSensor XXS 
measuring channel 1.

Carry out a span calibration, see Section 5.2 on 
Page 8.

203

The measured value of DrägerSensor XXS 
measuring channel 1 is in the negative range.

Carry out a fresh air calibration, see Section 5.2 on 
Page 8.

204

DrägerSensor XXS measuring channel 1 is faulty or 
not connected.

Check DrägerSensor XXS measuring channel 1, see 
Section 7.4 on Page 13.

205

Error during the bump test of DrägerSensor XXS 
measuring channel 1.

Repeat the bump test and calibrate or replace 
DrägerSensor XXS measuring channel 1 if 
necessary, see Section 7.4 on Page 13.

206

Faulty filter test.

Repeat the filter test with X-dock.

207

Faulty rise time test.

Repeat the rise test with X-dock.

208

User parameters of the sensor implausible.

Check and adjust the configuration of the sensor 
parameters.

212

Calibration interval of DrägerSensor XXS measuring 
channel 1 expired.

Carry out a calibration, see Section 5 on Page 8.

225

Compensation channel error on DrägerSensor XXS 
measuring channel 1.

Carry out a calibration, see Section 5 on Page 8.

Содержание Pac 6000

Страница 1: ...Dr ger Pac 6x00 8x00 Technical manual i...

Страница 2: ......

Страница 3: ...ration 7 5 Calibrating the gas detector 8 5 1 Opening the maintenance menu 8 5 2 Carrying out a manual fresh air calibration 8 5 3 Carrying out a manual span calibration 9 5 4 Carrying out manual cali...

Страница 4: ...and non earthed metal parts of the housing may store a combustible level of electrostatic charge Activities such as carrying the device in a bag or attached to a belt operating the button field or cle...

Страница 5: ...ormity are provided in the enclosed supplementary documentation order no 90 33 741 The name plate on the gas detector must not be concealed 1 Alarm LEDs 6 OK button 2 Operation signal D Light 7 button...

Страница 6: ...curacy After the gas detector has been switched on the current measured value is shown in the display Check whether the warning appears If it is displayed we recommend performing a bump test see Secti...

Страница 7: ...vely The information display closes after the device ID When the gas detector is on The following information can be displayed depending on the configuration Error codes Peak concentration Time weight...

Страница 8: ...the last confirmation with the OK button the entire password flashes 3 Press the OK button to confirm the password or the button to cancel the password 4 If the correct password is entered a signal to...

Страница 9: ...r the button to cancel the span calibration Span calibration passed A short double tone sounds and OK gas name is shown in an alternating display Press the OK button to return to measuring mode Span c...

Страница 10: ...Wrong software version Update the software 117 User parameters implausible Check and adjust the configuration of the user parameters 131 Gas detector faulty Contact the service of Dr ger 132 Paramete...

Страница 11: ...rror on Dr gerSensor XXS measuring channel 2 Carry out a calibration see Section 5 on Page 8 327 Gas detector faulty Contact the service of Dr ger 328 Compensation channel calibration error on Dr gerS...

Страница 12: ...n Page 8 351 Dr gerSensor XXS measuring channel 2 warming up warm up phase 1 Wait until warm up time is complete 352 Dr gerSensor XXS measuring channel 2 warming up warm up phase 2 Wait until warm up...

Страница 13: ...ing the Dr ger CC Vision Basic PC software 1 Use the communication module order no 83 18 587 to connect the gas detector to a PC or the Dr ger X dock maintenance station 2 The configuration can be car...

Страница 14: ...on Basic PC software as described Dr ger recommends signing on the sensor with the Dr ger CC Vision Basic PC software even if the sensor code is identical to the previous one 7 5 Replacing the battery...

Страница 15: ...iaphragm must be replaced due to the special diaphragm 8 Device settings Only trained and qualified personnel may change the device settings 8 1 Factory setting The factory settings may differ for cus...

Страница 16: ...Main alarm3 Instrument alarm 1 The intensity of the vibration depends on the temperature 2 After the initial battery pre alarm the battery s life span amounts to between 1 day and 2 weeks under normal...

Страница 17: ...centers as specified by the applicable regulations Ambient conditions during operation Temperature temperature depending on the sensor Up to 30 C 55 C briefly up to 1 h 40 C 55 C Humidity 10 90 r h n...

Страница 18: ...m Acknowledgeable Latching Alarm threshold A2 1 60 ppm 10 ppm 23 Vol 2 ppm Acknowledgeable Latching TWA threshold value 1 3 3 Not applicable for Pac 6000 30 ppm 10 ppm 1 ppm STEL threshold value 1 3 6...

Страница 19: ...1 100 ppm 0 2 ppm 20 ppm 50 ppm 1 ppm 1 Vol Acknowledgeable Latching TWA threshold value 1 20 ppm 0 1 ppm 1 9 ppm 25 ppm 0 5 ppm 0 5 Vol STEL threshold value 1 40 ppm 0 1 ppm 3 8 ppm 50 ppm 1 ppm 2 V...

Страница 20: ...A2 1 1 ppm 20 ppm 20 ppm 0 2 ppm 0 2 ppm Acknowledgeable Latching TWA threshold value 1 0 5 ppm 0 1 ppm 0 1 ppm STEL threshold value 1 0 5 ppm 0 1 ppm 0 1 ppm Mean value period 15 min 15 min 15 min Me...

Страница 21: ...60 ppm CO 23 O 2 60 CO 60 ppm Acknowledgeable O 2 CO Latching O 2 CO TWA threshold value 1 5 ppm H 2 S 30 ppm CO 30 ppm CO 30 ppm CO STEL threshold value 1 10 ppm H 2 S 60 ppm CO 60 ppm CO 60 ppm CO M...

Страница 22: ...2 No effect Ethane No effect No effect 0 2 Ethanol No effect No effect No effect Ethylene No effect No effect 1 Hydrogen 0 35 No effect 1 5 Hydrogen chloride No effect No effect No effect Hydrogen cy...

Страница 23: ......

Страница 24: ...ger Safety AG Co KGaA Revalstra e 1 23560 L beck Germany Tel 49 451 882 0 Fax 49 451 882 20 80 www draeger com 90 33 742 TH 4623 700 Dr ger Safety AG Co KGaA Edition 01 July 2017 Subject to alteratio...

Отзывы: