Technical data
21
en
10
.3
.1
Sensor
s for Pac 85
00 DUAL
H
2
S
LC /
CO
LC
O
2
/ CO
-LC
CO H
2
-CP
Measurin
g range
0
...
100
pp
m H
2
S
0
...
20
00
ppm CO
0.
..
25V
ol
%
O
2
0
...
200
0
p
pm C
O
0
...
20
00
pp
m
Cali
bration
g
as co
ncentration
5
...
90 ppm H
2
S
20
...
4
50
ppm CO
12
...
2
0
V
o
l% O
2
20
...
1
800
pp
m CO
10
00 pp
m H
2
20
...
1800
p
pm C
O
Te
mperature
ran
ge,
opera
tio
n
-40
...
50
°C
-4
0.
..
1
22°
F
-40
...
50
°C
-40
...
122
°F
-40
...
50
°C
-4
0.
..
1
22°
F
Alarm threshold
A1
1)
1)
Observe special
, custome
r-re
quest
ed se
tt
ing
s.
D
evi
ce
p
ar
amet
ers can
b
e
cha
nged
wit
h
th
e Dr
äger
CC-V
isi
on
Basic P
C
sof
tware
.
5 p
p
m H
2
S
30
ppm CO
19
O
2
30
CO
30
ppm
Ackno
wled
geab
le
O
2
:
-
CO:
La
tc
hi
ng
-
O
2
:
CO: -
-
Alarm threshold
A2
1)
10 ppm H
2
S
60
ppm CO
23
O
2
60
CO
60
ppm
Ackno
wled
geab
le
-
O
2
:
-
CO: -
-
La
tc
hi
ng
O
2
:
CO:
TW
A th
reshol
d
va
lue
1)
5 p
p
m H
2
S
30
ppm CO
-
30
p
pm C
O
-
30
ppm CO
STEL threshold value
1)
10 ppm H
2
S
60
ppm CO
-
60
p
pm C
O
-
60
ppm CO
Me
an
v
a
lu
e
pe
rio
d
15
min
15
m
in
15
min
Measureme
nt accu
ra
cy
Z
ero poi
nt:
H
2
S:
0.4
ppm
CO:
2p
pm
O
2
±0.4
V
o
l%
CO:
±2
pp
m
±6
p
pm
Sen
sitivity: [%
o
f th
e me
asured
va
lue]
H
2
S:
±5
CO:
±2
O
2
±1
CO
:
±2
±2
Lon
g-te
rm drif
t (20
°C)
Z
ero poi
nt:
H
2
S:
±0
.2
p
pm/a
CO:
±2
ppm/a
O
2
±0
.5 V
ol%/a
CO
:
±
2
ppm/a
±2
ppm/a
Sen
sitivity:
[%
of measured value/
year
]
H
2
S:
±5
CO:
±3
O
2
±1
CO
:
±3
±1/month
Sensor item numbe
r
2)
2)
T
he sensors h
a
ve a li
mite
d lif
e sp
an.
Excessive
stor
age p
e
riod
s
co
mpromise t
he
oper
ati
onal l
ife
time
of
th
e sensors.
The a
deq
ua
te
te
m
perat
ure
ran
ge f
or st
orag
e
is
0
..
.
35
°C
(
3
2
.
..
95
°F)
.
6
8
13
280
68
13
2
75
68
11
95
0
Sensor da
ta
she
et item numbe
r
90
33
5
11
90
33
5
10
9
0
23
924