Safety-related information

en

1

Safety-related information

1.1

Basic safety information

Before using this product, carefully read the associated 
instructions for use. This document does not replace the 
instructions for use.

Incorrect calibration

An incorrect calibration leads to incorrect measured values.

The sensitivity must be checked daily before first using the 
device, with a known concentration of the gas to be 
measured that corresponds to 25 to 50 % of the final 
concentration. The accuracy must amount to 0 to +20 % of 
the actual value. The accuracy can be corrected by 
calibration.

1.2

Safety information on explosion 
protection

Devices or components that are used in potentially explosive 
atmospheres and which are tested and approved in 
accordance with national, European or international explosion 
protection guidelines may only be used under the approved 
conditions in compliance with the statutory provisions.

Oxygen-enriched atmospheres

Explosion protection is not guaranteed in oxygen-enriched 
atmospheres (>21 Vol% O

2

).

Remove the device from the potentially explosive 
atmosphere.

Risk of explosion!

Do not open the gas detector in explosion-hazard areas.

Specific application conditions

In certain extreme conditions, open plastic parts and non-
earthed metal parts of the housing may store a combustible 
level of electrostatic charge.

Activities such as carrying the device in a bag or attached 
to a belt, operating the button field or cleaning the device 
with a damp cloth do not represent a significant 
electrostatic danger. However, if a mechanism which 
generates static charge, such as repeated rubbing against 
clothing, is identified, suitable precautionary measures 
must be implemented, e.g. the use of anti-static clothing 
and anti-static footwear.

2

Conventions used in this 
document

2.1

Definition of alert icons

The following alert messages are used in this document to 
provide and highlight areas of the associated text that require 
a greater awareness by the user. A definition of the meaning of 
each alert message is as follows:

WARNING

Indicates a potentially hazardous situation. If not avoided, it 
could result in death or serious injury.

NOTICE

Indicates a potentially hazardous situation. If not avoided, it 
could result in damage to the product or environment.

2.2

Typographical conventions

This symbol identifies information that make the product 
easier to use.

2.3

Glossary

Technical term

Explanation

Operation signal

A periodic optical (green LED) and/or 
acoustic signal.

D-Light

The D-Light allows the user to check and 
indicate compliance with certain settings 
(e.g. bump test interval). The green LED 
flashes for a short period and 
superimposes the optical operation 
signal.

Содержание Pac 6000

Страница 1: ...Dr ger Pac 6x00 8x00 Technical manual i...

Страница 2: ......

Страница 3: ...ration 7 5 Calibrating the gas detector 8 5 1 Opening the maintenance menu 8 5 2 Carrying out a manual fresh air calibration 8 5 3 Carrying out a manual span calibration 9 5 4 Carrying out manual cali...

Страница 4: ...and non earthed metal parts of the housing may store a combustible level of electrostatic charge Activities such as carrying the device in a bag or attached to a belt operating the button field or cle...

Страница 5: ...ormity are provided in the enclosed supplementary documentation order no 90 33 741 The name plate on the gas detector must not be concealed 1 Alarm LEDs 6 OK button 2 Operation signal D Light 7 button...

Страница 6: ...curacy After the gas detector has been switched on the current measured value is shown in the display Check whether the warning appears If it is displayed we recommend performing a bump test see Secti...

Страница 7: ...vely The information display closes after the device ID When the gas detector is on The following information can be displayed depending on the configuration Error codes Peak concentration Time weight...

Страница 8: ...the last confirmation with the OK button the entire password flashes 3 Press the OK button to confirm the password or the button to cancel the password 4 If the correct password is entered a signal to...

Страница 9: ...r the button to cancel the span calibration Span calibration passed A short double tone sounds and OK gas name is shown in an alternating display Press the OK button to return to measuring mode Span c...

Страница 10: ...Wrong software version Update the software 117 User parameters implausible Check and adjust the configuration of the user parameters 131 Gas detector faulty Contact the service of Dr ger 132 Paramete...

Страница 11: ...rror on Dr gerSensor XXS measuring channel 2 Carry out a calibration see Section 5 on Page 8 327 Gas detector faulty Contact the service of Dr ger 328 Compensation channel calibration error on Dr gerS...

Страница 12: ...n Page 8 351 Dr gerSensor XXS measuring channel 2 warming up warm up phase 1 Wait until warm up time is complete 352 Dr gerSensor XXS measuring channel 2 warming up warm up phase 2 Wait until warm up...

Страница 13: ...ing the Dr ger CC Vision Basic PC software 1 Use the communication module order no 83 18 587 to connect the gas detector to a PC or the Dr ger X dock maintenance station 2 The configuration can be car...

Страница 14: ...on Basic PC software as described Dr ger recommends signing on the sensor with the Dr ger CC Vision Basic PC software even if the sensor code is identical to the previous one 7 5 Replacing the battery...

Страница 15: ...iaphragm must be replaced due to the special diaphragm 8 Device settings Only trained and qualified personnel may change the device settings 8 1 Factory setting The factory settings may differ for cus...

Страница 16: ...Main alarm3 Instrument alarm 1 The intensity of the vibration depends on the temperature 2 After the initial battery pre alarm the battery s life span amounts to between 1 day and 2 weeks under normal...

Страница 17: ...centers as specified by the applicable regulations Ambient conditions during operation Temperature temperature depending on the sensor Up to 30 C 55 C briefly up to 1 h 40 C 55 C Humidity 10 90 r h n...

Страница 18: ...m Acknowledgeable Latching Alarm threshold A2 1 60 ppm 10 ppm 23 Vol 2 ppm Acknowledgeable Latching TWA threshold value 1 3 3 Not applicable for Pac 6000 30 ppm 10 ppm 1 ppm STEL threshold value 1 3 6...

Страница 19: ...1 100 ppm 0 2 ppm 20 ppm 50 ppm 1 ppm 1 Vol Acknowledgeable Latching TWA threshold value 1 20 ppm 0 1 ppm 1 9 ppm 25 ppm 0 5 ppm 0 5 Vol STEL threshold value 1 40 ppm 0 1 ppm 3 8 ppm 50 ppm 1 ppm 2 V...

Страница 20: ...A2 1 1 ppm 20 ppm 20 ppm 0 2 ppm 0 2 ppm Acknowledgeable Latching TWA threshold value 1 0 5 ppm 0 1 ppm 0 1 ppm STEL threshold value 1 0 5 ppm 0 1 ppm 0 1 ppm Mean value period 15 min 15 min 15 min Me...

Страница 21: ...60 ppm CO 23 O 2 60 CO 60 ppm Acknowledgeable O 2 CO Latching O 2 CO TWA threshold value 1 5 ppm H 2 S 30 ppm CO 30 ppm CO 30 ppm CO STEL threshold value 1 10 ppm H 2 S 60 ppm CO 60 ppm CO 60 ppm CO M...

Страница 22: ...2 No effect Ethane No effect No effect 0 2 Ethanol No effect No effect No effect Ethylene No effect No effect 1 Hydrogen 0 35 No effect 1 5 Hydrogen chloride No effect No effect No effect Hydrogen cy...

Страница 23: ......

Страница 24: ...ger Safety AG Co KGaA Revalstra e 1 23560 L beck Germany Tel 49 451 882 0 Fax 49 451 882 20 80 www draeger com 90 33 742 TH 4623 700 Dr ger Safety AG Co KGaA Edition 01 July 2017 Subject to alteratio...

Отзывы: