background image

6

Use

en

4

Use

4.1

Preparations for use

4.1.1

Initial start-up

The gas detector remains in deep sleep mode upon delivery 
and must be activated during the initial start-up.
1. Hold down the [

] button for approx. 3 s. 

This activates the gas detector.

4.1.2

Switching on the gas detector

1. Hold down the [OK] button for approx. 3 s.

The following is displayed or activated:

Display elements, LEDs, alarm signal and vibration alarm

Self-test

Software version and gas name

Alarm thresholds A1 and A2

Time to next calibration (configurable)

Time to expiration of the bump test interval (configurable)

Before every use, check whether the display elements and 
information are displayed correctly.

A warm-up phase takes place when first switching on the gas 
detector (duration depends on the sensor type; see the sensor 
data sheet).

4.1.3

Switching off the gas detector

Hold down both keys for approx. 3 s, until shut-down is 
complete. 

4.2

Before entering the workplace

WARNING

Serious damage to health!
An incorrect calibration can lead to incorrect measurement 
results, which may result in serious damage to health.

Before performing safety measurements, check the 
calibration by way of a bump test, adjust as necessary, 
and check all alarm elements. If national regulations exist, 
the bump test must be performed in accordance with 
these regulations.

WARNING

Incorrect measurement results!
The gas inlet opening is fitted with a filter that protects against 
dust and water. Contamination may change the qualities of the 
dust and water filter.

Do not damage the filter. Immediately replace damaged or 
blocked filters.

To ensure correct function:

Do no cover the gas inlet opening.

Place the gas detector on clothing near the mouth.

At temperatures below -20 °C, there may be deviations of 
>10 % of the measured value if the corresponding sensor 
was calibrated at room temperature. Dräger recommends 
calibration at the primary application temperature if 
measurements will be taken at very low temperatures. This 
enables the highest possible measurement accuracy.

After the gas detector has been switched on, the current 
measured value is shown in the display.
Check whether the warning [!] appears. If it is displayed, we 
recommend performing a bump test, see Section 4.2.3 on 
Page 6.

4.2.1

Activating the quick menu

The quick menu can be activated using the Dräger CC-Vision 
Basic PC software. 

Up to two preferred functions can be saved in the quick menu. 
The following functions are available:

Fresh air adjustment

Bump test

4.2.2

Opening the quick menu

Prerequisite:

The quick menu is activated.

To open the quick menu:
1. Press the [

] button three times within three seconds. A 

double signal tone sounds.

2. Depending on the configuration, the fresh air calibration 

symbol or the bump test signal will flash.

3. Press the [OK] button to start the feature or press the [

button to switch to the next feature or to return to 
measuring mode (depending on the configuration).

The quick menu closes automatically after 60 s of inactivity.

4.2.3

Carrying out a manual bump test

WARNING

Danger to health! Do not inhale the test gas.

Pay attention to the hazard information in the respective 
safety data sheets.

Two modes can be selected for a bump test. Use the Dräger 
CC-Vision Basic PC software to set the mode.

Quick bump test (test for alarm triggering)

Checks whether the pre-alarm threshold (A1) is 
exceeded (or not reached in the case of O

2

).

Checks whether the concentration remains above the 
alarm for a specified period of time.

Checks whether the test duration remains below a 
specified maximum time period.

Advanced bump test (test for accuracy)

Checks whether the defined test gas concentration is 
reached within a specified tolerance. The tolerance is 
defined with a default value for each gas in the gas 
detector, but can be adjusted by the user as needed.

Checks whether the test gas concentration remains 
within the tolerance window for a specified time period.

Checks whether the test duration remains below a 
specified maximum time period.

The bump test can be performed as follows:

Manual bump test (with quick menu)

Bump test with X-dock (see instructions for use for the 
Dräger X-dock)

Bump test with the Dräger Bump Test Station (see the quick 
reference guide on the Dräger Bump Test Station)

Содержание Pac 6000

Страница 1: ...Dr ger Pac 6x00 8x00 Technical manual i...

Страница 2: ......

Страница 3: ...ration 7 5 Calibrating the gas detector 8 5 1 Opening the maintenance menu 8 5 2 Carrying out a manual fresh air calibration 8 5 3 Carrying out a manual span calibration 9 5 4 Carrying out manual cali...

Страница 4: ...and non earthed metal parts of the housing may store a combustible level of electrostatic charge Activities such as carrying the device in a bag or attached to a belt operating the button field or cle...

Страница 5: ...ormity are provided in the enclosed supplementary documentation order no 90 33 741 The name plate on the gas detector must not be concealed 1 Alarm LEDs 6 OK button 2 Operation signal D Light 7 button...

Страница 6: ...curacy After the gas detector has been switched on the current measured value is shown in the display Check whether the warning appears If it is displayed we recommend performing a bump test see Secti...

Страница 7: ...vely The information display closes after the device ID When the gas detector is on The following information can be displayed depending on the configuration Error codes Peak concentration Time weight...

Страница 8: ...the last confirmation with the OK button the entire password flashes 3 Press the OK button to confirm the password or the button to cancel the password 4 If the correct password is entered a signal to...

Страница 9: ...r the button to cancel the span calibration Span calibration passed A short double tone sounds and OK gas name is shown in an alternating display Press the OK button to return to measuring mode Span c...

Страница 10: ...Wrong software version Update the software 117 User parameters implausible Check and adjust the configuration of the user parameters 131 Gas detector faulty Contact the service of Dr ger 132 Paramete...

Страница 11: ...rror on Dr gerSensor XXS measuring channel 2 Carry out a calibration see Section 5 on Page 8 327 Gas detector faulty Contact the service of Dr ger 328 Compensation channel calibration error on Dr gerS...

Страница 12: ...n Page 8 351 Dr gerSensor XXS measuring channel 2 warming up warm up phase 1 Wait until warm up time is complete 352 Dr gerSensor XXS measuring channel 2 warming up warm up phase 2 Wait until warm up...

Страница 13: ...ing the Dr ger CC Vision Basic PC software 1 Use the communication module order no 83 18 587 to connect the gas detector to a PC or the Dr ger X dock maintenance station 2 The configuration can be car...

Страница 14: ...on Basic PC software as described Dr ger recommends signing on the sensor with the Dr ger CC Vision Basic PC software even if the sensor code is identical to the previous one 7 5 Replacing the battery...

Страница 15: ...iaphragm must be replaced due to the special diaphragm 8 Device settings Only trained and qualified personnel may change the device settings 8 1 Factory setting The factory settings may differ for cus...

Страница 16: ...Main alarm3 Instrument alarm 1 The intensity of the vibration depends on the temperature 2 After the initial battery pre alarm the battery s life span amounts to between 1 day and 2 weeks under normal...

Страница 17: ...centers as specified by the applicable regulations Ambient conditions during operation Temperature temperature depending on the sensor Up to 30 C 55 C briefly up to 1 h 40 C 55 C Humidity 10 90 r h n...

Страница 18: ...m Acknowledgeable Latching Alarm threshold A2 1 60 ppm 10 ppm 23 Vol 2 ppm Acknowledgeable Latching TWA threshold value 1 3 3 Not applicable for Pac 6000 30 ppm 10 ppm 1 ppm STEL threshold value 1 3 6...

Страница 19: ...1 100 ppm 0 2 ppm 20 ppm 50 ppm 1 ppm 1 Vol Acknowledgeable Latching TWA threshold value 1 20 ppm 0 1 ppm 1 9 ppm 25 ppm 0 5 ppm 0 5 Vol STEL threshold value 1 40 ppm 0 1 ppm 3 8 ppm 50 ppm 1 ppm 2 V...

Страница 20: ...A2 1 1 ppm 20 ppm 20 ppm 0 2 ppm 0 2 ppm Acknowledgeable Latching TWA threshold value 1 0 5 ppm 0 1 ppm 0 1 ppm STEL threshold value 1 0 5 ppm 0 1 ppm 0 1 ppm Mean value period 15 min 15 min 15 min Me...

Страница 21: ...60 ppm CO 23 O 2 60 CO 60 ppm Acknowledgeable O 2 CO Latching O 2 CO TWA threshold value 1 5 ppm H 2 S 30 ppm CO 30 ppm CO 30 ppm CO STEL threshold value 1 10 ppm H 2 S 60 ppm CO 60 ppm CO 60 ppm CO M...

Страница 22: ...2 No effect Ethane No effect No effect 0 2 Ethanol No effect No effect No effect Ethylene No effect No effect 1 Hydrogen 0 35 No effect 1 5 Hydrogen chloride No effect No effect No effect Hydrogen cy...

Страница 23: ......

Страница 24: ...ger Safety AG Co KGaA Revalstra e 1 23560 L beck Germany Tel 49 451 882 0 Fax 49 451 882 20 80 www draeger com 90 33 742 TH 4623 700 Dr ger Safety AG Co KGaA Edition 01 July 2017 Subject to alteratio...

Отзывы: