Supplement – Infinity
®
Delta Series Software VF10.1
81
Supplement for Infinity Delta Series Software
Reprocessing list
Reprocessing procedures
Validated reprocessing procedures
At the time of product-specific validation, the following reprocessing procedures showed good material
compatibility and effectiveness:
The effectiveness of the listed reprocessing procedures has been validated by independent laboratories
that are certified to the standard ISO 17025.
Disinfectants
Use disinfectants that are nationally approved and are suitable for the particular reprocessing procedure.
Surface disinfectant
At the time of the test, the surface disinfectants listed in the following table showed good material
compatibility. They can be used in addition to the surface disinfectants listed in the section "Validated
reprocessing procedures".
The manufacturers of the surface disinfectants have verified at least the following spectra of activity:
– Bactericidal
– Yeasticidal
– Virucidal or virucidal against enveloped viruses
Observe the specifications of the surface disinfectant manufacturers.
Components
Surface
disinfection
with cleaning
Manual
cleaning
followed by
disinfection
by immersion
Machine
cleaning with
thermal
disinfection
Steam
sterilization
Special
reprocessing
measures
Delta series
monitors
Yes
N/A
N/A
N/A
N/A
Procedure
Agent
Manufacturer
Concen-
tration
Contact
time
Tempera-
ture
Surface disinfection with
cleaning
Dismozon Plus
BODE Chemie
1.6 %
15 min
N/A
Oxycide
Ecolab USA
2.34 %
5 min
N/A