Supplement for Infinity Delta Series Software
Inspection
Perform inspections at regular intervals and observe the following specifications.
CNAP / Cuff Controller
Visual Inspection
Perform a visual inspection before every use and in accordance with your hospital's policy.
1
Make sure that the housing is not cracked or broken and there are no signs of spilled liquids or
damage.
2
Inspect all accessories (for example, sensors and cables). Do not use if there are any signs of damage.
3
Examine all system cables and power plugs and discontinue use if there are any signs of damage.
4
Inspect all patient cables, leads, and strain reliefs for general condition. Make sure the connectors are
properly engaged at each end.
Inspection/ Safety Check
Inspection and safety checks of devices must be performed according to suggested intervals.
Scope of inspection/safety checks for the CNAP Pod/ Cuff Controller
Safety checks are no substitute for preventive maintenance measures (including preventive replacement
of wear parts) as identified by the manufacturer. Perform safety checks at the indicated intervals.
1
Check accompanying documents:
– Instructions for use are available
2
Perform a functional test of the following features according to the instructions for use:
– CNAP pod
– Cuff controller
Checks
Interval
Responsible personnel
Inspection/safety checks Every 2 years
Service personnel
Metrological checks
Every 2 years
Service personnel
WARNING
Risk of medical device failure
If safety checks are not performed on a regular basis, the proper operation of the medical device
can be compromised. Perform safety checks at the indicated interval.