Supplement – Infinity
®
Delta Series Software VF10.1
47
Supplement for Infinity Delta Series Software
Quick Reference Table -
etCO2
* Setup
Click on the
etCO2*
parameter box to execute
etCO2*
setup functions:
Calibration
Check
Checks the calibration of the CO2 sensor with a test filter. The monitor reports
whether the calibration check has passed or failed.
NOTE
The sensor must be removed from the cuvette and zeroed before performing a
calibration check. During the calibration check the sensor must be connected to the
test filter.
CO2 Zero
(Only available if
a Dräger
Mainstream CO
2
module is
connected.)
Zeroes the CO2
sensor. The CO2
sensor stores a new zero point for CO2
measurements.
NOTE
The sensor must be removed from the cuvette before zeroing. The sensor is zeroed
in room air. Do not breathe on the cuvette during zeroing.
Apnea Alarm
Displays the
Apnea Alarm
menu.
– RRc Apnea Time
– Apnea Archive
etCO2 Alarm
Accesses the
Alarm Limits
table where you can
set upper and lower alarm thresholds. See the
Alarms chapter for more information about
setting and displaying alarm limits.
WARNING
etCO2 alarms do not activate until the
second breath is detected after turning on
the monitor, admitting a patient or
disconnecting and reconnecting the
Mainstream CO2 sensor or module.
Not applicable
Menu Item
Description
Settings
Menu Item
Description
Settings
Source
Describes what CO2*
device is connected and
displays parameters on the monitor.
NOTE
– If no CO2* device is connected, “
None
”
displays (read-only).
– If one CO2* device is connected to the
monitor, the setting automatically displays
the appropriate device (read-only).
– If two CO2
*
devices are connected, the
settings are made available to the user to
select the desired CO2* device.
(See also the “Multiple CO2 Devices” section.)
– SCIO
– Mainstream
– None