Supplement – Infinity
®
Delta Series Software VF10.1
21
Supplement for Infinity Delta Series Software
Alarm History
Table
The monitor stores up to 50 physiological alarm events for each patient (the oldest log entry is overwritten).
Events are deleted when the patient is discharged. Data is stored in the monitor and remains with the
patient during Pick and Go transports. Data also survives power shutdowns. The
Alarm History
table
records all high-priority and medium-priority alarms as well as selected low-priority alarms for MultiGas
and Dräger Mainstream CO
2
monitoring (e.g., alarms relating to module disconnection). The
Alarm
History
table also records every activation and deactivation of cardiac bypass mode, every change of
patient category, and records whenever you perform an
All Alarms OFF
/
All Alarms Paused
or
Audio
Paused
/
Alarm Silence
. Switching the device on and off is not recorded.
Alarm Latching
Some alarms are latched: they continue to annunciate visually and audibly until you acknowledge them
manually, even if the condition that caused the alarm no longer exists. Other alarms may be latched only
partially, as indicated in the following table:
To acknowledge (or silence) a latched alarm, press the
Alarm Silence
fixed key.
NOTE
High-priority and medium-priority alarms do not
latch in OR mode.
Alarm Latching Behavior
High-priority alarms are latched (visual and auditory alarm signals continue).
Medium-priority alarms only latch the alarm message; it continues to display when the alarm condition
ceases, while the parameter box stops flashing and the alarm tone ceases.
Low-priority alarms cease as soon as the cause of the alarm disappears.