Supplement for Infinity Delta Series Software
20
Supplement – Infinity
®
Delta Series Software VF10.1
Auto Set Alarm Limits
Alarm Limits
Table
Function
Description
Available Settings
Auto Set
Sets alarm limits based on current values:
Not applicable
NOTE
:
• The monitor recalculates the upper
and lower alarm limits based on the
parameter values in the
Current
column.
•
Auto Set
applies to all displayed
parameters and ST parameters
only.
• If a calculated limit value falls
outside the range for that
parameter, the parameter’s alarm
limits remain unchanged.
Parameters
Upper Limit
Lower Limit
Ta, T1a, T1b,
T2a, T2b, T3a,
T3b
< 107% of
current value
< 93% of
current value
ΔT1, ΔT2, ΔT3,
PVC/min,
inCO2/inCO2*
No change
No change
SpO2/SpO2*
Adults 100
Neonates 98
Current value
–
(value x 5%)
ΔSpO2%
Current value
+20%
None
ST
Current value
+2.0 mm
Current value
–
2.0 mm
MultiGas Agent < 105% of
current value
< 95% of
current value
BIS
No change
No change
TOF-Cnt
No change
No change
PCCO
Current value
+30%
Current value
–
30%
PCCI
Current value
+30%
Current value
–
30%
MultiGas O2
100%
21%
tp
< 125% of
current value
< 80% of
current value
All others
Current value
+25%
Current value
–
20%