Supplement for Infinity CentralStation Wide
12
Supplement
–
Infinity CentralStation Wide
–
VG2.1.2
Parameter order
The following illustration shows the
Parameter order
dialog, which lets the user configure the order of
parameters and how they display on the
Main screen
.
To access the
Parameter order
dialog:
1
Select
System setup...
on the main menu bar.
2
Select the
Parameter order
tab.
3
Next to
Template
,
in the pull-down box
,
select from 25 predefined templates.
4
In the
Name
field, enter a name for the template you selected in step 3.
5
Next to
Display mode
, select:
Auto
–
Unavailable waveforms and parameters from the source device will not be displayed.
Manual
–
Unavailable waveforms and parameters from the source device will display blank
Only the first 11 parameters from the source device are available for display
6
Under
Rearrange parameter order
, use the up and down arrows to organize the waveforms under
Waveform
, and parameters under
Parameter box
. Multiple parameter boxes can be selected and
moved at one time, using drag and drop functionality.
7
Click
Apply
to save the selections.
NOTE
Only named templates are available in the
Main
screen layout
tab or in the customize layout
dialog.
NOTE
If a template is active on the
Main screen
, the
template name can be modified but cannot be
deleted.
NOTE
Select the
System default
button to reset to the
factory default settings.