INFINITY CENTRAL STATION & M300 QUICK REFERENCE GUIDE
|
23
THIS GUIDE IS INTENDED FOR REFERENCE ONLY. REFER TO INSTRUCTIONS FOR USE MANUAL.ADDITIONAL INFORMATION
tient on the ICS Main screen.
2.
Select the
Alarms
button in the menu bar of the patient’s
BedView page.
3.
Select the
Arrhythmia
tab.
4.
If desired, select the desired arrhythmia mode using the
ARR
mode
buttons (
Basic
or
Advanced
or
OFF
). OFF requires clinical
password.
5.
Select the corresponding button in the Alarm column to select
the alarm priority (
High, Medium, Low
, or
OFF
). When an ar-
rhythmia alarm is turned OFF the following symbol appears:
Asystole (
ASY)
and ventricular fibrillation (
VF
) events
are always assigned a high alarm priority. These priorities can-
not be changed (appear grey).
6.
Select the corresponding setup field in the
Rate
column to set
the rate, if applicable.
7.
Select the corresponding setup button in the Count column to
set the count, if applicable.
8.
Use one of the following settings in the
Archive
column to de-
termine what happens in response to an alarm:
o
Off
: no event is stored and no recording is generated.
o
Store
: stores the event in the disclosure database.
o
Record
: generates a strip recording.
o
Store/Rec
: stores the event in the disclosure database
and generates a strip recording.
9.
Repeat steps 5 through 8 for each parameter to be configured.
10.
Select the
Apply
button.