INFINITY CENTRAL STATION & M300 QUICK REFERENCE GUIDE
|
12
THIS GUIDE IS INTENDED FOR REFERENCE ONLY. REFER TO INSTRUCTIONS FOR USE MANUAL.ADDITIONAL INFORMATION
BedView Menu Bar
Key
Function
Bed audio pause
Pauses all active alarm tones for the selected
patient (the duration of the audio pause depends on
the setting of the bedside monitor).
Audio Pause
Suppresses any alarms for the patient for a pre-
configured period.
Alarms...
Accesses the setup page for activating the alarm func-
tion for individual parameters and setting alarm limits.
Record
Starts a timed strip recording.
Mark event
(M300 only)
Enables you to save and store a manual event to the
disclosed database, where it can then be viewed from
the events tab.
For the M300, this is accomplished by selecting the
Mark
event button from the Bed/View menu bar. From
a device other than the M300, marking an event must
take place at that specific device (which will be sent to
the ICS).
Print screen
Prints the contents of the current screen on a
connected laser printer.
ECG report
(non-diag,)
Request a non-diagnostic ECG report.
ECG relearn
Relearns the normal ECG waveform template for
processing arrhythmias and ST.
Resp relearn
Relearns the breath detection threshold (for bedside
monitors only).
ECG views...
Enables you to access All leads or ST complexes
pages.
Trends/Data
Accesses the
Trends graph, Trends table, Full disclo-
sure
, and
Events
pages, from which you can view
graphical and tabular trend data, Full disclosure data,
and Events data for a specific patient.