Instructions for use

  |  

Dräger BG ProAir

  

Description 

Status messages

Statuses of the Connect ECU

– On mode

The Connect ECU switches from On mode to Standby 

mode in the following cases:
– If the apparatus has not been moved for an extended 

period of time and the system pressure is 0 bar

– If both buttons on the Connect ECU are pressed for at 

least 2 seconds and the system pressure is below 

6 bar

– Standby mode

When the apparatus is not being moved, the 

Connect ECU switches from Standby mode to Eco mode 

after 60 minutes.

The Connect ECU switches from Standby mode to On 

mode in the following cases:
– When the compressed oxygen cylinder is opened
– When a button is pushed on the Connect ECU for at 

least 2 seconds

– When the function key is removed

– When the user breathes from the apparatus. When the 

user is no longer breathing from the apparatus, the 

Connect ECU returns to Standby mode after approx. 

10 seconds.

Even when the Connect ECU is in Standby mode, it 

recognises when the user is breathing from the 

apparatus. If the compressed oxygen cylinder is 

closed or empty, the no-oxygen alarm will be triggered.

– Eco  mode

As soon as the closed-circuit breathing apparatus is 

moved, the Connect ECU switches back from Eco mode 

to Standby mode.

The Connect ECU switches from Eco mode to On mode in 

the following cases:
– When a button is pushed on the Connect ECU
– When the function key is removed

– Off

When the batteries are empty, the Connect ECU is 

switched off. When new batteries are inserted, the 

Connect ECU automatically switches to On mode. An 

acoustic signal sounds. Various information and queries 

are shown on the display.

In Eco mode, opening the cylinder will not be recognised 

and the no-oxygen alarm is deactivated.

The various states of the Connect ECU allow for maximum 

operational readiness with minimum power consumption.

Details on the display when turning the apparatus on

The following information is shown when the compressed 

oxygen cylinder is opened. Other information and queries are 

shown if the Connect ECU is turned on differently 

(pushbutton, insertion of batteries, removal of function key). 

The following information and queries are displayed in 

succession when the Connect ECU switches to On mode:
– Displays the battery charge status
– Queries whether the high-pressure leak test is to be 

carried out

– Display of filling pressure for the compressed oxygen 

cylinder and the "Remaining time until low-pressure 

alarm 1"

Alternatively, pushing the button on the left or right displays 

either the "Elapsed time" or the "Temperature". 

Use as distress signal unit

The user can trigger a distress signal using the manual alarm 

button if the user cannot retreat from the hazard area.

The Connect ECU can automatically issue an alarm if the 

user is no longer moving. To do so, the function key must be 

removed from the Connect ECU. If no further motion is 

detected, a pre-alarm sounds after a defined time (default 

setting: 30 seconds). If any movement is detected within the 

following approx. 15 seconds, the pre-alarm is cancelled. If no 

motion is detected, the main alarm sounds.

Main alarm (motion sensor)

The alarm is triggered 

15 seconds after the pre-

alarm (motion sensor), once 

the pre-alarm (motion sen-

sor) is not acknowledged.

Main alarm (manual)

The alarm is triggered when 

the manual alarm button is 

pressed.

Battery alarm

The alarm is triggered when 

the charge status of the bat-

tery reaches a critical level 

during operation.

1) The value is configurable

Time remaining until low-

pressure alarm 1

The electronic monitoring 

system uses the cylinder 

pressure and the respiration 

to compute the time until the 

warning whistle sounds. 

A respiration of 30 L/min is 

initially assumed. The com-

putation of the remaining time 

is then updated once per sec-

ond with the specific respira-

tion (but at least 30 L/min).

Elapsed time

The elapsed time indicates 

how long the Connect ECU 

has been switched on.

Temperature

The temperature shows the 

ambient temperature of the 

Connect ECU.

Содержание BG ProAir

Страница 1: ...Instructions for use Dr ger BG ProAir More languages available for download at draeger com ifu Dr ger BG ProAir...

Страница 2: ......

Страница 3: ...200 0 02 bar h elapsed time 1 4 2 3 27 27 11 16 9 10 8 29 28 24 25 26 14 15 12 13 19 20 23 22 21 5 6 7 17 18 31 30 8 7 6 9 1 2 3 4 5...

Страница 4: ...digitalt exemplar p andra spr k i databasen f r teknisk doku mentation www draeger com ifu SV 9300712 ET Selle kasutusjuhendi saab teistes keeltes elektroonili sel kujul alla laadida tehnilise dokume...

Страница 5: ...use 12 4 3 1 Freezing the ice packs for apparatuses with ice cooler 12 4 3 2 Inserting ice packs for apparatuses with ice cooler 12 4 3 3 Donning the closed circuit breathing apparatus 12 4 3 4 Activ...

Страница 6: ...to potential hazards The meanings of the warning notices are defined as follows 2 2 Trademarks The following website lists the countries in which the trademarks from Dr ger are registered www draeger...

Страница 7: ...g cooler options are available Ice cooler The ice cooler contains ice packs that can be refrozen after use Regeneration cooler The regeneration cooler PCM phase chage material cooler contains a latent...

Страница 8: ...following information is shown when the compressed oxygen cylinder is opened Other information and queries are shown if the Connect ECU is turned on differently pushbutton insertion of batteries remo...

Страница 9: ...mmersion of the apparatus in water e g when navigating a waterhole in a mine will not impair the function of the closed circuit breathing apparatus If the closed circuit breathing apparatus has been c...

Страница 10: ...ion on approvals see supplement 9300233 3 6 Type identifying markings and explanation of symbols 3 6 1 Type identifying marking The type identifying marking of the BG ProAir Class 4P is located on the...

Страница 11: ...ored for 6 months The apparatus must have successfully passed the leak test and be sealed with the protective cap The air tight shrink wrapped or installed CO2 absorber was not transported on a vehicl...

Страница 12: ...ing aid to prevent the ice packs from deforming Make sure that no ice is deposited on the ice packs If necessary place ice packs in a bag seal the bag and freeze the ice packs whilst they remain in th...

Страница 13: ...g the automatic distress signal function Remove the function key from the Connect ECU This activates the motion sensor The symbol is displayed 4 4 During use 4 4 1 General information The chemical pro...

Страница 14: ...ecess when tightening the bayonet ring Rotate the bayonet ring clockwise onto the connection to fasten the breathing hose The breathing hose is correctly fitted when a clicking sound is heard while ti...

Страница 15: ...k the catches on the fasteners under the tabs see Figure B 1 Push the fasteners closed see Figure B 2 Connecting piece The direction of flow is marked with an arrow on the connecting piece The directi...

Страница 16: ...g circuit may contain base which may leak when disassembling the closed circuit breathing apparatus and lead to chemical burns Avoid skin contact Wear protective gloves and goggles In case of eye cont...

Страница 17: ...th release levers 1 Swivel the spring bridge 90 and remove it from the closed circuit breathing apparatus 2 Release the locking mechanisms between the CO2 absorber and breathing bag as well as between...

Страница 18: ...4 Prepare the CO2 absorber Dispose of the disposable absorber For the refill absorber remove the soda lime and the two non woven mats and discard them The sieve can be reused after disinfection 5 2 3...

Страница 19: ...rmation 5 2 5 Assembling the closed circuit breathing apparatus 5 2 5 1 Overview 1 Install the CO2 absorber see Installing the CO2 absorber page 19 2 Install the breathing air cooler see Installing th...

Страница 20: ...nect the red medium pressure hose to the breathing air cooler The quick connect coupling must engage When connecting make sure that a click is heard If necessary pull on the hose and check whether the...

Страница 21: ...th the locking mechanisms to the CO2 absorber and breathing air cooler Every locking mechanism has two lock buttons with engagement hooks Ensure the locking mechanisms on the CO2 absorber and breathin...

Страница 22: ...en the Connect ECU on the shoulder strap 5 2 6 Testing the closed circuit breathing apparatus 5 2 6 1 Notes for testing The following test sequence applies for both apparatuses with constant oxygen do...

Страница 23: ...he selector switch on the test unit to 2 Tightly pinch the exhalation hose with your hand Use the face spanner from the BG ProAir Test Set if necessary 3 Pump once The test unit must display at least...

Страница 24: ...ically turns on and runs a self test 3 Briefly press the bypass button Oxygen must be heard to flow into the circuit flow noise 5 2 6 13 Checking the minimum valve 1 Set the selector switch on the tes...

Страница 25: ...b 400 when handling the CO2 absorber WARNING Reduced period of use possible If the expiration date of the soda lime is exceeded the intended period of use for the apparatus may not be reached Note the...

Страница 26: ...s are shown on the display 7 Confirm the battery type 8 Turn off the Connect ECU 6 Transport Condensate forms in the breathing circuit during use After use transport the closed circuit breathing appar...

Страница 27: ...5 L min Bypass 80 L min Minimum valve 80 L min Temperature during continu ous use1 1 according to EU type examination 6 C to 30 C Air pressure 900 hPa to 1200 hPa Relative humidity 0 100 Storage temp...

Страница 28: ...ylinder 2 200 O2 CGA USA R360395 BG ProAir 2 200 O2 Configurator R360560 Accessories Dr ger FPS 7000 RP R56326 Dr ger FPS COM 5000 R62700 klar pilot anti fogging spray R56542 BG ProAir disposable abso...

Страница 29: ...tightness test with positive pressure Lower pressure to 7 0 5 mbar Press OK button Pressure difference 1 mbar min Engage lever in spring bridge Pump to activate Opens between 4 and 8 mbar Relief valve...

Страница 30: ...e pressure Lower pressure to 7 0 5 mbar Press OK button Pressure difference 1 mbar min Relief valve Engage lever in spring bridge Pump to activate Opens between 4 and 8 mbar Bypass Open cylinder Brief...

Страница 31: ...Annex Instructions for use Dr ger BG ProAir 31...

Страница 32: ......

Страница 33: ......

Страница 34: ...9300708H Manufacturer Dr ger Safety AG Co KGaA Revalstra e 1 D 23560 L beck Germany 49 451 8 82 0 9300708 1114 800 en Dr ger Safety AG Co KGaA Edition 01 2021 10 Subject to alterations www draeger com...

Отзывы: