Dräger BG ProAir

  

Maintenance 

5 Maintenance

5.1 Maintenance 

intervals

Dräger recommends the following maintenance intervals. 

Observe any national guidelines that may stipulate different 

intervals.

5.2 Maintenance 

work

5.2.1  General notes on handling

Breathing hoses

The breathing hoses are secured on the connecting piece, on 

the CO

2

 absorber and on the breathing air cooler with 

bayonet rings.
– To remove, rotate the bayonet rings anti-clockwise and 

use the flap to pull the breathing hoses off the relevant 

component.

– Every breathing hose has one thick, narrow flap and one 

thin, wide flap. The flaps, connecting piece, CO

2

 absorber 

and breathing air cooler are each marked with a triangle or 

a circle. The end with the narrow flap (marked with a 

triangle) is secured to the connecting piece on the side 

with the exhalation valve (marked with a triangle) and to 

the breathing air cooler (marked with a triangle). The end 

with the wide flap (marked with a circle) is secured to the 

connecting piece on the side with the inhalation valve 

(marked with a circle) and to the CO

2

 absorber (marked 

with a circle). When fitting the items, make sure that the 

width of the flap corresponds to the width of the recess 

and that the markings match.

Push the end of the breathing hose onto the 

connection. The flap must be positioned in the recess 

when tightening the bayonet ring.

Rotate the bayonet ring clockwise onto the connection 

to fasten the breathing hose. The breathing hose is 

correctly fitted when a clicking sound is heard while 

tightening and the bayonet ring has some play. The 

bayonet ring (1) must sit flush with the edge of the 

relevant connection.

Medium-pressure hoses

The medium-pressure hoses are connected to the CO

2

 

absorber and to the breathing air cooler with quick connect 

couplings (see overview graphic at start of instructions, items 

5, 6, 7). The quick connect couplings can be released from 

the CO

2

 absorber and breathing air cooler without using tools 

and are self-locking.

If a pneumatic hose is inserted into a connector multiple 

times, it must be verified that the pneumatic hose can 

continue being used.

After use

Every 6

m

onths

Annually

Eve

ry 5

yea

rs

E

ve

ry1

0y

e

ar

s

Visual inspection of all parts,

Testing the functional integrity and 

tightness of the closed-circuit 

breathing apparatus

X

X

1)

1) Applies for stored apparatuses.

Cleaning and disinfecting the 

closed-circuit breathing apparatus

X

X

1)

Replacing the CO

2

 absorber

X

X

2)

2) Applies for CO

2

 absorbers that are installed in a closed-circuit breath-

ing apparatus and whose functional integrity and tightness were 
tested with the apparatus. The protective cap must be attached to the 
connecting piece of the closed-circuit breathing apparatus.

Hydrostatic testing of the com-
pressed oxygen cylinder

3)

3) by a notified body

X

Major overhaul

4)

4) only performed by expert personnel

X

44939

45771

1

1

Содержание BG ProAir

Страница 1: ...Instructions for use Dr ger BG ProAir More languages available for download at draeger com ifu Dr ger BG ProAir...

Страница 2: ......

Страница 3: ...200 0 02 bar h elapsed time 1 4 2 3 27 27 11 16 9 10 8 29 28 24 25 26 14 15 12 13 19 20 23 22 21 5 6 7 17 18 31 30 8 7 6 9 1 2 3 4 5...

Страница 4: ...digitalt exemplar p andra spr k i databasen f r teknisk doku mentation www draeger com ifu SV 9300712 ET Selle kasutusjuhendi saab teistes keeltes elektroonili sel kujul alla laadida tehnilise dokume...

Страница 5: ...use 12 4 3 1 Freezing the ice packs for apparatuses with ice cooler 12 4 3 2 Inserting ice packs for apparatuses with ice cooler 12 4 3 3 Donning the closed circuit breathing apparatus 12 4 3 4 Activ...

Страница 6: ...to potential hazards The meanings of the warning notices are defined as follows 2 2 Trademarks The following website lists the countries in which the trademarks from Dr ger are registered www draeger...

Страница 7: ...g cooler options are available Ice cooler The ice cooler contains ice packs that can be refrozen after use Regeneration cooler The regeneration cooler PCM phase chage material cooler contains a latent...

Страница 8: ...following information is shown when the compressed oxygen cylinder is opened Other information and queries are shown if the Connect ECU is turned on differently pushbutton insertion of batteries remo...

Страница 9: ...mmersion of the apparatus in water e g when navigating a waterhole in a mine will not impair the function of the closed circuit breathing apparatus If the closed circuit breathing apparatus has been c...

Страница 10: ...ion on approvals see supplement 9300233 3 6 Type identifying markings and explanation of symbols 3 6 1 Type identifying marking The type identifying marking of the BG ProAir Class 4P is located on the...

Страница 11: ...ored for 6 months The apparatus must have successfully passed the leak test and be sealed with the protective cap The air tight shrink wrapped or installed CO2 absorber was not transported on a vehicl...

Страница 12: ...ing aid to prevent the ice packs from deforming Make sure that no ice is deposited on the ice packs If necessary place ice packs in a bag seal the bag and freeze the ice packs whilst they remain in th...

Страница 13: ...g the automatic distress signal function Remove the function key from the Connect ECU This activates the motion sensor The symbol is displayed 4 4 During use 4 4 1 General information The chemical pro...

Страница 14: ...ecess when tightening the bayonet ring Rotate the bayonet ring clockwise onto the connection to fasten the breathing hose The breathing hose is correctly fitted when a clicking sound is heard while ti...

Страница 15: ...k the catches on the fasteners under the tabs see Figure B 1 Push the fasteners closed see Figure B 2 Connecting piece The direction of flow is marked with an arrow on the connecting piece The directi...

Страница 16: ...g circuit may contain base which may leak when disassembling the closed circuit breathing apparatus and lead to chemical burns Avoid skin contact Wear protective gloves and goggles In case of eye cont...

Страница 17: ...th release levers 1 Swivel the spring bridge 90 and remove it from the closed circuit breathing apparatus 2 Release the locking mechanisms between the CO2 absorber and breathing bag as well as between...

Страница 18: ...4 Prepare the CO2 absorber Dispose of the disposable absorber For the refill absorber remove the soda lime and the two non woven mats and discard them The sieve can be reused after disinfection 5 2 3...

Страница 19: ...rmation 5 2 5 Assembling the closed circuit breathing apparatus 5 2 5 1 Overview 1 Install the CO2 absorber see Installing the CO2 absorber page 19 2 Install the breathing air cooler see Installing th...

Страница 20: ...nect the red medium pressure hose to the breathing air cooler The quick connect coupling must engage When connecting make sure that a click is heard If necessary pull on the hose and check whether the...

Страница 21: ...th the locking mechanisms to the CO2 absorber and breathing air cooler Every locking mechanism has two lock buttons with engagement hooks Ensure the locking mechanisms on the CO2 absorber and breathin...

Страница 22: ...en the Connect ECU on the shoulder strap 5 2 6 Testing the closed circuit breathing apparatus 5 2 6 1 Notes for testing The following test sequence applies for both apparatuses with constant oxygen do...

Страница 23: ...he selector switch on the test unit to 2 Tightly pinch the exhalation hose with your hand Use the face spanner from the BG ProAir Test Set if necessary 3 Pump once The test unit must display at least...

Страница 24: ...ically turns on and runs a self test 3 Briefly press the bypass button Oxygen must be heard to flow into the circuit flow noise 5 2 6 13 Checking the minimum valve 1 Set the selector switch on the tes...

Страница 25: ...b 400 when handling the CO2 absorber WARNING Reduced period of use possible If the expiration date of the soda lime is exceeded the intended period of use for the apparatus may not be reached Note the...

Страница 26: ...s are shown on the display 7 Confirm the battery type 8 Turn off the Connect ECU 6 Transport Condensate forms in the breathing circuit during use After use transport the closed circuit breathing appar...

Страница 27: ...5 L min Bypass 80 L min Minimum valve 80 L min Temperature during continu ous use1 1 according to EU type examination 6 C to 30 C Air pressure 900 hPa to 1200 hPa Relative humidity 0 100 Storage temp...

Страница 28: ...ylinder 2 200 O2 CGA USA R360395 BG ProAir 2 200 O2 Configurator R360560 Accessories Dr ger FPS 7000 RP R56326 Dr ger FPS COM 5000 R62700 klar pilot anti fogging spray R56542 BG ProAir disposable abso...

Страница 29: ...tightness test with positive pressure Lower pressure to 7 0 5 mbar Press OK button Pressure difference 1 mbar min Engage lever in spring bridge Pump to activate Opens between 4 and 8 mbar Relief valve...

Страница 30: ...e pressure Lower pressure to 7 0 5 mbar Press OK button Pressure difference 1 mbar min Relief valve Engage lever in spring bridge Pump to activate Opens between 4 and 8 mbar Bypass Open cylinder Brief...

Страница 31: ...Annex Instructions for use Dr ger BG ProAir 31...

Страница 32: ......

Страница 33: ......

Страница 34: ...9300708H Manufacturer Dr ger Safety AG Co KGaA Revalstra e 1 D 23560 L beck Germany 49 451 8 82 0 9300708 1114 800 en Dr ger Safety AG Co KGaA Edition 01 2021 10 Subject to alterations www draeger com...

Отзывы: